AQUATERR INSTRUMENTS EC-300 User Manual Download Page 10

 

10 

CALIBRATING THE SALINITY FUNCTION (Salinity Multimeter) 

 

Caution: This should only be done as a last resort after you have tried everything else and 

you are certain that it is required. 

 

We have provided two methods of calibrating the salinity function. The first method is more 
convenient for laboratory or controlled conditions; the second is more convenient in the field.  
 
The calibration knob “CAL” is located on the upper left hand side of the meter’s panel. Make 
sure that the knob to the right of the meter’s panel “W” is turned the furthest clockwise 
position for liquids and set to 100 for soils. 
 
First Method: 
 

Using a Bench Top Meter

 – If you have a calibrated bench-top or laboratory EC meter:  

1.  Prepare a water sample at or near the field EC using table salt and your bench top 

meter. 

2.  Follow instructions for measuring liquid salinity. Remember to give the meter two to 

three minutes to reach temperature equilibrium.   

3.  Compare the Multimeter meter reading to the bench top reading. 
4.  Adjust the Multimeter calibration knob “CAL” (left of the meter) until the meters 

agree. 

 
 
Second Method: 
 

Air

 – This will set your meter to a repeatable value, but not to the accuracy which can be 

obtained by calibration with a bench top meter. 

1.  Place the Multimeter in the sample and wait two or three minutes to reach 

temperature equilibrium (the longer the better). 

2.  Quickly take the meter out of the sample, wipe off the tip and press EC button. 
3.  When the EC-ON light comes on adjust the calibration knob “CAL” (left of the 

meter) until the Multimeter reads 0.1 in the air. This is the best starting point for 
field use. 

 
If neither of these methods work, remove the battery cover (four small Phillips head screws), 
and carefully disconnect the battery cable from the battery. Then locate the coarse zero set 
(see diagram), and using a small screwdriver, rotate the small screw counterclockwise to 
increase, clockwise to decrease, and follow the above directions.  
 

 
 
 
 
 
 
 

Summary of Contents for EC-300

Page 1: ...AQUATERR I N S T R U M E N T S M T EC 300 User s Manual Aquaterr Digital Soil Moisture Temperature and Salinity Meters ...

Page 2: ...2 AQUATERR INSTRUMENTS Leader in Portable Soil Measuring Technology ...

Page 3: ...ovel and then insert the probe If the ground is still too hard use an auger or a pipe to create the hole Do not hit the meter end This may result in loss of calibration or destruction of your meter Do not cut the roots of plants if avoidable Insert the probe with the flat edges parallel to the roots Test the soil under plants by inserting the probe at an angle Be especially careful of the tap root...

Page 4: ...bly soft and moist area to start 2 Gently insert the probe into the ground six inches is enough 3 Press the MSTR or TEST button on the front panel The meter will respond with the soil moisture reading 4 Press the TEMP button and the meter will show you the soil temperature 5 To read EC Salinity Meter only set the moisture calibration knob W upper right corner to the moisture reading from step 3 No...

Page 5: ... all times If you have very soft water EC below 200 ppm or 30 ms call technical service for calibration recommendations 2 While the probe is in the water push the MSTR button the only push button on the Moisture Meters While pushing down on the MSTR button turn the SET knob until the meter reads 100 clockwise increases the meter reading 3 The Moisture Meter is now ready to use MEASURING SOIL MOIST...

Page 6: ...t each location Make sure that the sensor has reached temperature equilibrium in the soil for two to three minutes before calibrating and recalibrate when temperature changes occur 1 Push the probe into the soil to the desired depth If the ground is too hard use an auger The sensing window and wings must be in contact with the soil for repeatability 2 Push the MSTR button The reading is the soil m...

Page 7: ...emperature sensor is covered This means covering the entire winged section 2 Wait until the sensor stabilizes typically 2 3 minutes and the sensor s temperature approaches the soil s temperature Generally the more moisture in the soil the faster the sensor stabilizes 3 Push the TEMP button The meter reading indicates the sensor temperature 4 WIPE OFF the probe between readings COMMOM PROBLEMS Soil...

Page 8: ...ensitive plants irrigate in the upper one half of the zone Use the lower end of the zone for mature and drought resistant plants Yellow Stress is apt to occur although most plants will survive if water condition is temporary Orange Damaging stress to most plants Red Permanent wilting and death is likely While the color coding should only be used as a reference it points out that plant stress not o...

Page 9: ...ill precipitate in the soil s free available water and will saturate out in this soil water mixture There is not enough water present for the salts to mix with the water evenly For this reason a sample of the soil must be taken then the moisture and salt content of the soil sample has to be determined and the results can be used to calibrate the EC readings of the EC 300 Salinity Multimeter PROBE ...

Page 10: ...er two to three minutes to reach temperature equilibrium 3 Compare the Multimeter meter reading to the bench top reading 4 Adjust the Multimeter calibration knob CAL left of the meter until the meters agree Second Method Air This will set your meter to a repeatable value but not to the accuracy which can be obtained by calibration with a bench top meter 1 Place the Multimeter in the sample and wai...

Page 11: ...inless Steel OVERALL LENGTH 36 Inches approx 91 44cm TEMPERATURE RANGE 4 44 to 65 55 deg C METER ACCURACY 1 of full scale POWER SOURCE 9 volt battery LIFE OF POWER SOURCE Approx 3 months WEIGHT Approx 5 pounds 2 26Kg SCALE MOISTURE 0 100 TEMPERATURE 0 199 deg F 0 92 77 deg C EC 2000 µS microSiemens ...

Reviews: