DX-300 – DIVER HELD MAGNETOMETER

 User Manual 

 

 

N.B.  THE  CURRENT  TYPE  OF  CHARGERS  SUPPLIED  (JAN  2011 

ONWARDS) PROVIDE A CONSTANT RATE OF CHARGE FOR UP TO 14 
HOURS;  FOLLOWED  BY  A  TRICKLE  CHARGE  (MAINTENANCE) 
MODE.  WHILST  CHARGING  AT  THE  REGULAR  RATE  THE  LED 
LIGHT WILL GLOW  ORANGE, ONCE THE CHARGING PERIOD BEEN 
COMPLETED  THE  CHARGER  DROPS  INTO  “TOP  UP”  CHARGE 
MODE,  AT  THIS  STAGE  THE  LIGHT  CHANGES  TO  GREEN  WITH  A 
YELLOW FLASH. 

 
 

During  subsequent  partial  recharge/top  up  requirements;  batteries  will  not  be 
damaged by being left on charge for a longer period than necessary, this is due 
to the controlled charge current being automatically followed by a trickle charge 
mode.  If the batteries become fully discharged - due to natural extended usage 
or  through  the  unit  accidentally  being  left  in  the  on  state  –  the  unit  should  be 
fully  recharged  as  soon  as  possible  to  avoid  irreparable  damage  to  the  battery 
cells. 

 

N.B.  DO  NOT  USE  ANY  BATTERY  CHARGER  OTHER  THAN  THAT 

SUPPLIED; AS THIS COULD RESULT IN DAMAGE TO THE BATTERY 
PACK.

 

 
 

When  the  charge  period  has  elapsed,  disconnect  the  AC  supply  to  the  charger 

first

,  and  then  unplug  the  connector  from  the  socket  on  the  detector.  Charging 

can be terminated before the 14 hour period has elapsed if the unit just requires a 
top  up  charge.  NB.  The  operational  usage  time  on  a  full  charge  with  a  healthy 
battery  pack  will  be  in  the  order  of  24  hours,  recharging  will  require  a  1  hour 
charge time for each 2 hours of usage; in other words a 2:1 usage/charge time to 
replenish.  Some  earlier  types  of  charger  had  a  6  hour  time  out  period  requiring 
the  charger  to  be  switched  off  and  back  on  to  extend  the  charging  to  periods 
exceeding the 6 hour cut off. 

 
2.  For initial testing on land, you will have to find a location where you can position 

yourself at least 20 metres from any large ferrous objects in all directions.  Get rid 
of anything containing ferrous/magnetic items on your body or in your clothing – 
and leave your mobile/cell phone well out of range.   An empty farm field, park, 
sports field or beach is ideal for testing.  If possible park your car at the edge of 
the  field  (at  least  20  metres  from  where  you  will  be  testing),  or  find  a  location 
where there is some large ferrous target in one direction only.        

 
3.  To prepare the DX-300 for use, plug the connector on the Sensor Probe into the 

SENSOR socket and screw down the retaining collar.  This should be good finger 
tightness only.  

 
 

N.B.PLIERS OR OTHER TOOLS SHOULD NOT BE USED TO TIGHTEN THE 

PLUG COLLARS.  THESE ARE DESIGNED TO BE USED FINGER TIGHT 
WITH  THE  BLACK  RUBBER  BULKHEAD  SEAL  IN  PLACE  ON  THE 
PLUG  PINS.    OVER-TIGHTENING,  OR  INSTALLING  THE  PLUG 

Summary of Contents for DX-300

Page 1: ...Aquascan Int Ltd all rights reserved This document must not be replicated in whole or part without the prior express permission of Aquascan International Ltd Issue date 24th June 2014 AQUASCAN INTERNATIONAL LTD AQUASCAN HOUSE HILL STREET NEWPORT SOUTH WALES UK NP20 1LZ Tel 44 1633 841117 Fax 44 1633 254829 Email info aquascan co uk Website www aquascan co uk ...

Page 2: ... Charging the batteries 8 Underwater Operation 9 Care and Maintenance 10 Battery Care Checking Battery Performance Status 12 Troubleshooting Guide No Response from the DX 300 Module Erratic Performance from the DX 300 Module 14 DX 300 Specifications 15 DX 300 Accessories 17 DX 300 Optional Accessories 18 DX 300 Professional Kit 19 Manufacturer Contact Information 20 EC Declaration of Conformity Pr...

Page 3: ...f the pipe or cable E O D Security There are unlimited applications in all types of underwater investigation for the presence of ordinance weapons etc Salvage The DX 300 is a useful tool for the successful recovery of lost items either in low visibility or where lost in soft sediment It is the ideal tool to complement surface operated searching The tool most requested by users of boat towed magnet...

Page 4: ...ial areas or under transformers and high tension power lines will generally give the user problems with false readings In the water it is just important to be clear of any influence that may be generated by a large steel vessel and of course equipment on the diver such as a steel dive cylinder If you are searching for small items of debris such as a cannonball a pistol a hand tool you will probabl...

Page 5: ...ity of the DX 300 and how to use the controls It is much easier to learn to operate the unit without the challenges of limited visibility restrictive dive equipment currents etc So please take the time to perform this Initial Testing of your unit before venturing into the water with it 1 Charging the batteries The DX 300 is supplied only partially charged and will require a full 14 hours charging ...

Page 6: ...has elapsed if the unit just requires a top up charge NB The operational usage time on a full charge with a healthy battery pack will be in the order of 24 hours recharging will require a 1 hour charge time for each 2 hours of usage in other words a 2 1 usage charge time to replenish Some earlier types of charger had a 6 hour time out period requiring the charger to be switched off and back on to ...

Page 7: ...s so just start turning one way until you either reach the audio threshold or the stop point of the control If you reach the stop point without finding the audio threshold start turning the other way until you do Once you have found the audio threshold the first time you test the unit the SET AUDIO control should be in the near vicinity of the correct location for future uses so you will normally ...

Page 8: ...d in a number of ways the choice will be influenced by the conditions and diver preference The most obvious choice is worn clipped to any suitable location on the diver with the D Ring mounting system provided with the unit Other options include the quick release belt harness strapping it to the upper arm or mounting it to the DX 300 Armsaver optional accessory supplied with the Professional syste...

Page 9: ...lugs make sure that the unit is switched off and wash down with clean fresh water NEVER use chemicals to clean your DX 300 as these will damage o rings and seals which can cause them to fail and allow the incursion of water into the control unit 2 Remove the Sensor Probe and Bonephone Headphones plugs from the bulkhead sockets and make sure the Rubber Seal is free from sand or other debris remove ...

Page 10: ...formance it is important to charge your DX 300 periodically when not in use Checking Battery Pack Performance Note For routine maintenance this should be carried out every three months The recommended procedure is as follows 1 0 Connect the Sensor Probe to your DX 300 module 1 1 Power on your DX 300 and leave it on overnight to fully discharge the battery 1 2 Power off the DX 300 and disconnect th...

Page 11: ...age will result in total lack of functionality 2 Check that both Sensor Probe and Bonephone are correctly connected 3 Try using an alternative Bonephone AQ1B Headphones 4 Try using an alternative Sensor Probe if available 5 Check that both the Sensor Probe and Bonephone are in good condition Check by substitution if possible or by monitoring the nominal resistance of Bonephone Headphones Connectin...

Page 12: ...ne are clean If necessary reduce the contact resistance by slightly prising the gap in the male contacts this will increase the firmness of the connector on insertion 4 Check for visible damage to Sensor Probe cable and connector which could have caused water ingress 5 Check that the male connector seals are in place and clean 6 If you are unable to determine the cause of the erratic behaviour con...

Page 13: ...mm Probe Length 750mm CABLE Construction Twin core shield Diameter 6 7mm Sheath Polyurethane Length 2m approx 6ft POWER Power Internal Rechargeable 2200 NiMH battery Charger Universal 100 250v AC Charge time from empty 14Hrs CONTROLS Power On Off Incorporated with Sensitivity Sensitivity Switch Off Low Med High Extra High Set Audio Multi turn Pot WEIGHTS Control Unit 1Kg Probe Cable 0 72Kg Basic o...

Page 14: ...Headphones feature an adjustable volume control and a comfortable fit for the user AQ 4 062 DX 300 Bonephone with extending flex cable This version is in all other respects the same as the AQ 4 061 The option of a flexible cable that will extend and retract is preferred by some divers particularly in low visibility conditions to minimise the risk of the cable fouling any obstructions AQ 4 031 DX 3...

Page 15: ...h 100v to 250v AC 50 60HZ input voltage range this unit is used to recharge the internal battery pack of the DX 300 Control Module The charger can be supplied with option of US UK or European type Power Connector in the form of a push fit shoe or as a conventional power lead DX 300 Control unit ...

Page 16: ...TOMETER User Manual 16 DX 300 OPTIONAL ACCESSORIES Optional DX 300 Armsaver probe extended Optional DX 300 Armsaver probe retracted DX 300 Control Unit mounted in optional Armsaver Close up of DX 300 Armsaver Probe connection ...

Page 17: ... Manual 17 DX 300 Professional Kit Includes Transit Storage case for DX 300 Control Unit Sensor Probe D Ring Waist Belt Charger Bonephone Underwater Headphones and Armsaver kit DX 300 PROFESSIONAL KIT DX 300 Professional Kit Case closed ...

Page 18: ...equipment or to arrange a repair please use the following contact information Mailing Address Aquascan International Limited Aquascan House Hill Street Newport South Wales NP20 1LZ United Kingdom Tel 44 0 1633 841117 Fax 44 0 1633 254829 E mail Sales Enquiries sales aquascan co uk General Inquiries info aquascan co uk Technical Support support aquascan co uk Website www aquascan co uk ...

Page 19: ...claration will lose its authority Equipment description Diver Held Magnetometer Model DX 300 Manufacturer Aquascan International Ltd Aquascan House Hill Street Newport NP20 1LZ S Wales Tel 44 0 1633 841117 Fax 44 0 1633 254829 Email info aquascan co uk Applicable directives EN 55011 General Emission Standard Class A Group 1 EN 50082 Generic Immunity Standard Part 2 A technical construction file fo...

Reviews: