Microfreeze User’s Guide
10
F
IGURE
1:
O
RIENTATION OF THE SENSOR RELATIVE TO THE CONVEYOR BELT AND PRODUCT MOTION AXIS
.
✓
Please foresee free space (minimum 50 cm) around the top measurement unit (TMU) to facilitate the
installation of the product-thickness sensor at the antenna side as indicated in Figure 2. The product
thickness sensor is attached to a stainless-steel base plate which is mounted to the two wall-mount
spacers of the HF receiver unit. All units need to be mounted on metal frame which needs to be
accessible from the back to fix the different units.
✓
Mount the computational unit in the vicinity of the TMU and BMU with the cable glands pointed towards
the floor. Please foresee a save and easy access to the TMU, BMU and the computational unit to
facilitate the maintenance on the MICROFREEZE measurement system.
F
IGURE
2:
S
TAINLESS STEEL BASE PLATE FOR MOUNTING THE SENSOR MODULE AND THICKNESS SENSOR
.
✓
The conveyor belt thickness at the electromagnetic-axis should be constant (toothed sections of the
conveyor belt should not be placed in the EM field-of-view).
✓
The HF units and the computational unit should be mounted rigidly on a frame limiting any vibrations
as much as possible.
✓
Make sure there is no electromagnetic coupling from other devices operating in the same HF frequency
range: 2.4-2.5 GHz and 5.725-5.875 GHz. Respect a distance of 1 m from the MICROFREEZE system.