24
5.7
Description of menu items
As they appear in the menu structure
Possible parameter units are described on page 44
Main Menu
Mass Tot
totalizer of mass flow in selected units
Mass Subtot
subtotalizer of mass flow
Mass Flow
mass flow
Totalizer
main totalizer of volume flow
Subtot
subtotalizer of volume flow
Flow rate
flow rate (volume compensated when Comp on is set, Vc is displayed)
Temperature
displays oil temperature in selected unit
Tot d Vc
main totalizer of compensated volume flow for distillate fuel
Subtot d Vc
subtotalizer of compensated volume flow for distillate fuel
Tot H Vc
main totalizer of compensated volume flow for heavy fuel
Subtot H Vc
subtotalizer of compensated volume flow for heavy fuel
Tot L Vc
main totalizer of compensated volume flow for Lube oil
Subtot L Vc
subtotalizer of compensated volume flow for Lube oil
Setup
displays all setup parameters (edit with user, service code)
Display test
Test of all display segments
Setup Menu
enter from Main Menu “SETUP”
Unit vol
select units for volume display
Unit ti
select units for time display
Unit °t
select units for temperature display
Unit m
select units for mass display
Dn*
select nominal size of the hydraulics
→
pre set during
calibration
Vc*
displays the volume of the measuring chamber
→
set after calibration
Trip res
select if trips (all subtotals) can be reset (yes, no)
LFC
select Low Flow Cut off (no counting below this value, displays zero [0])
Comp OFF
compensation is turned off (volume and mass)
Comp On
compensation is turned on
(if mass is off, it will only compensate to Norm-Volume)
Mass OFF
Mass calculation is turned off
Mass On
Mass calculation is turned on (outputs are in mass)
Oil Fuel
select usage type Fuel oil
°t Limit
enter the thermal threshold for distillate and heavy fuel oil
dd 880kg/m
3
enter the density of currently used distillate fuel
dH 990 kg/m
3
enter the density of currently used heavy fuel
Oil Lube
select usage type Lube oil if counter is used to measure Lube oil
dL 900kg/m
3
enter the density of current Lube oil
o1 / o2 / AnA
Outputs, see details on next page
noErrors
no error messages
Errors
error messages displayed (see
Troubleshooting
for error message details)
noAlarms
no alarm messages
Alarms
alarm messages displayed (see
Troubleshooting
for alarm message
details)
Summary of Contents for CONTOIL VZF II
Page 9: ...9...