Aqualytic AL10pH Instruction Manual Download Page 5

5

AL10pH

4-2 Equipment Required
 

i)   Combination pH electrode

 

ii)  pH 4.0 & pH 7.0 buffer solutions

4-3 Two Point Calibration
 

i)   Connect the pH electrode to the BNC socket (Figure 1, 5) and immerse  

 

 

in pH 7.0 buffer solution.

 

ii)  Switch the instrument on by pressing the Power ON Button (Figure 1, 2).

 

iii)  Adjust „pH 7 Adjust „ (Figure 1, 6) until the display reads exactly pH 7.0.

 

iv)  Rinse the electrode in distilled water.

 

v)  Place the electrode into buffer pH 4.0 solution.

 

vi)  Adjust „pH 4 Adjust „ (Figure 1, 7) until the display reads exactly pH 4.0.

 

vii) Rinse the electrode in distilled water.

 

viii) Repeat steps iii) to vii) at least twice.

4-4 Single Point Calibration
 

i)  Connect the pH electrode to the BNC socket and immerse in a standard  

 

 

buffer solution (eg pH 7.0 or higher if available).

 

ii)  Switch the instrument on by pressing the Power ON Button (Figure 1, 2).

 

iii)  Adjust the “pH 7 Adjust „ (Figure 1, 7) until the display reads the same value as the  

 

 

standard buffer solution used.

5.  TAKING MEASUREMENTS

After calibration, the unit is ready for taking measurements.
  i)  Connect the pH electrode to the BNC socket.
  ii)  Switch the instrument on by pressing the Power ON Button.
  iii)  Immerse the electrode in the solution being measured; the instrument will display the  
    pH value. 
  iv)  During measurement, pressing the Data Hold Button (Figure 1, 4) will freeze the  
   

display.   At the same time the display will show a Data Hold indicator.  To release 

     the display freeze function, press the Data. Hold Button again; the Data Hold  
    indicator will disappear from the display.
  v)  After taking a measurement, rinse the electrode in distilled water.
   

Summary of Contents for AL10pH

Page 1: ...GB AL10pH pH Instruction Manual...

Page 2: ...ROL EDUCATION SWIMMING POOLS WATER CONDITIONING High quality compact unit that is designed for easy operation Water resistant front panel with easy to read LCD display and rubberised function keys Unp...

Page 3: ...D 21 5 mm 0 7 digit height Measurement Range 0 14 pH Resolution 0 01 pH Accuracy 0 07 pH pH 5 pH 9 0 1 pH pH 4 pH 10 0 2 pH pH 1 pH 3 9 pH 10 1 pH 13 Main instrument only 23 5 C after calibration Inpu...

Page 4: ...rotective cover 4 CALIBRATION 4 1 Overview The meter is supplied with calibration by mV signal that simulates an ideal pH electrode mV output 0 mV at pH 7 based on 25 C working temperature Ho wever be...

Page 5: ...socket and immerse in a standard buffer solution eg pH 7 0 or higher if available ii Switch the instrument on by pressing the Power ON Button Figure 1 2 iii Adjustthe pH7Adjust Figure1 7 untilthedisp...

Page 6: ...n pH 7 25 C red 90 mL traceable to N I S T 721249 pH Buffer Solution pH 10 25 C red 90 mL traceable to N I S T 8 IMPORTANT INFORMATION FOR pH ELECTRODES General information Non refillable gel electrod...

Page 7: ...solution please rinse the electrode with distilled water tap water for removal of possible residues of salt and to mini mize carryover effects 2 In case of air bubbles remove them by shaking 3 During...

Page 8: ...ision Aqualytic Schleefstra e 8 12 44287 Dortmund Tel 49 0 2 31 9 45 10 755 Fax 49 0 2 31 9 45 10 750 sales aqualytic de www aqualytic de Germany Technical changes without notice Printed in Germany 04...

Reviews: