-21-
7.16 4mA Adjustment
If the 4-20 mA setting is turned ON (7.2
Setting the 4-20 mA Output
), the following two
menus will appear. The first menu outputs a constant 4 mA while allowing for a small
amount of adjustment. The adjustment can be made using the
and
buttons.
This
adjustment will allow the operator to make the S200 TUr agree with a PLC or SCADA
system. The adjustment limits are ± 200 counts or about ± 0.2 mA.
This setting will be slightly different on each instrument as each S200 TUr will be factory
set to 4.00mA. Press the
button when adjustments are complete to save this setting and
move on to the 20mA adjustment.
7.17 20mA Adjustment
This menu operates similar to the previous menu. This menu outputs a constant 20 mA
while allowing for a small amount of adjustment. The adjustment can be made using the
and
buttons. The adjustment limits are ± 1000 counts or about ± 1 mA.
This setting will be slightly different on each instrument as each S200 TUr will be factory
set to 20.00mA.
7.18 Saving Configuration Settings
If extended settings are set to
OFF
, pressing the
button will save all settings and the
S200 TUr will automatically return to the normal
AUTO
mode of the instrument.
If extended settings are set to
On
, after the last menu of the extended settings, pressing the
button will save all settings and the S200 TUr will automatically return to the normal
AUTO
mode of the instrument.
The
CONFIG
menu may be used at any time to reset or change any of the parameters.
The
CONFIG
menu may be exited at any point in the menu by using the
MODE/EXIT
key. Any features that have been modified will be saved.