3.4.2 Pump calibration
Once the priming procedure is complete, the quantity of ml (milliliters) per stroke that the pump is able
to supply must be defined, this information is necessary for a correct dosage of the chemical and
essential if you want to use the PPM function described in the "PUMP SETUP" paragraph.
ATTENTION: in order to use this function, the pump MUST be primed.
To calibrate the pump, follow the steps below:
1. Make sure that the pump has
been installed on the plant and
it has been primed using the
function
described
in
the
previous paragraph (§ 3.4.1)
2. Connect the suction tube
(complete with foot filter) into a
BEKER-type
test
tube
graduated in ml (milliliters);
3. Fill the test tube until reaching
the maximum limit with the
chemical product that will be
used
during
the
normal
operation of the pump;
4. Power up the pump, enter the
SETTINGS
>
GENERAL
MENU, follow the instructions
on the display: enter the
maximum value of the beker by
turning the encoder knob, then
press the knob to confirm;
5. Set the number of strokes the
pump will perform;
6. Once the button has been
pressed to confirm, the pump
will start for the number of
strokes set;
7. At the end, check how much
chemical is left in the beker and
write down this value on the
display using the knob, finally
press to confirm this value;
8. the next screen will show the
value calculated by the pump in
ml per stroke.
9. To exit the calibration
function,
keep
the
button
pressed for 2 seconds.
Summary of Contents for HC151+ CI(J)
Page 2: ...HC151 MULTI HC200 MULTI ENGLISH ...
Page 20: ...HC151 MULTI HC200 MULTI ENGLISH Example ...
Page 28: ...HC151 MULTI HC200 MULTI ENGLISH EXPLODED VIEWS HC151 EXPLODED VIEW ...
Page 30: ...HC151 MULTI HC200 MULTI ADSP9500073 13 03 2020 Rev 1 0 30 36 HC200 EXPLODED VIEW ...
Page 32: ...HC151 MULTI HC200 MULTI ADSP9500073 13 03 2020 Rev 1 0 32 36 PUMP HEAD EXPLODED VIEW ...