Extension cables and alarm cables
Item
Length
Approval
Type
Item no.
Supply cable Control Unit Sweden
2.0 metres
CEE 7/VII
Y003-B/B
CCUS
Supply cable Control Unit USA/Canada
2.4 metres
UL817, CSA-
TA-3/2,4mG/SVT
CCUU
CC22.2 n 21
Extension cable kit Heater Jacket
6
2.5 metres
CE, UL, CSA
Silflex, 3G1,0
HJEC-2M
Extension cable kit Heater Jacket
6
5.0 metres
CE, UL, CSA
Silflex, 3G1,0
HJEC-5M
Extension cable kit Heater Jacket
6
10.0 metres
CE, UL, CSA
Silflex, 3G1,0
HJEC-10M
Alarm Cable Heater Jacket
2.5 metres
CE, UL, CSA
Silflex, 3G1,0
HJAC-2M
Alarm Cable Heater Jacket
5.0 metres
CE, UL, CSA
Silflex, 3G1,0
HJAC-5M
Alarm Cable Heater Jacket
10.0 metres
CE, UL, CSA
Silflex, 3G1,0
HJAC-10M
11.
Fault tracing
Problem
Probable cause
Corrective action
7
HJ does not become warm
Power switch not pressed in on
the HJCU
No voltage
Blown fuse
Faulty cable connection HJ
Short circuit in the HJ
Faulty cable connection Temp
HJCU or HJ faulty
Press the power switch on the
HJCU
Check the incoming voltage
Change the main fuse in the HJCU
(F1)
Check the connection
Contact the dealer
Check the connection
Contact the dealer
HJ becomes significantly
warmer than the set value
Faulty temp element
Faulty overheating cut-out
Faulty HJCU
Contact the dealer
Contact the dealer
Contact the dealer
12.
Dealer
Dealer:
AQ M-Tech AB
Adress:
Bolandsgatan 10
SE-753 23 Uppsala
Sweden
Phone:
+46 (0)18-470 29 00
Fax:
+46 (0)18-10 50 04
Web:
E-mail:
6. Cable kit includes two cables (Temp & Heater cables).
7.
The Heater Jacket Control Unit (HJCU) must be disconnected from the main supply when working inside the enclosure.