Note: Regardless the setting for the recording interval, the values “Ah” and “Wh” are only
calculated by the unit once per second. When using an interval setting of < 1 s,
several identical values of Ah and Wh are written into the CSV.
6.10.6
Possible Reasons for Battery Test Stop
The battery test function can be stopped in different ways:
Manual stop on the front panel using the STOP button.
After the max. test time has been reached and action “End of test” was set for it.
After the max. battery capacity to be consumed has been reached and action “End
of test” was set for it.
Any device alarm which would also switch off the DC input, like OT.
Passing the threshold UDV (discharge voltage), which is equivalent to any voltage
drop on the DC input caused by whatever reason.
Note: After an automatic stop, caused by any of the listed reasons, the test cannot be
continued or run again immediately. The full battery configuration has to be run
from the start. This can be done using the BACK touch key.