Chapter 5
– Grounding and
Cabling the IONizer
Apprion Proprietary and Confidential
Page 13
Chapter 5
Grounding and Cabling the IONizer
Overview
This chapter provides the cabling procedures for the IONizer 4000.
Grounding
All IONizers must be properly grounded before making power and signal connections. Apprion
recommends a UL Listed #10 AWG wire that is suitable for this purpose, and a ring tongue terminal,
Panduit P/N P10-8R or equivalent. The terminal is to be crimped to the wire using the correct crimping
tool as recommended by the terminal manufacturer. The torque rating on the nut is 8 to 10 inch
pounds. The wire should be kept as short as possible while using grounding practices that are compliant
with local codes and practices.
Note:
Ensure that the connection to a proper earth ground is made by certified and authorized
personnel. The ground must conform to all applicable codes and regulations. The materials required to
connect to a proper earth ground are defined by local conditions and must be procured locally to ensure
that the correct safety environment is achieved.
Grounding Procedure
The grounding stud is located on the bottom of the IONizer at the location shown:
Bottom view of IONizer 4020
Attach the earth ground wire (not supplied) to the ring terminal attached to the IONizer’s grounding
stud. Ensure that the ring terminal is seated against the IONizer’s metal case.
Summary of Contents for IONizer 4000 Series
Page 6: ......
Page 8: ......
Page 12: ......
Page 29: ...Appendix A Approved Antennas Apprion Proprietary and Confidential Page 21 Vertical H Plane...
Page 30: ...Appendix A Approved Antennas Page 22 Apprion Proprietary Confidential Horizontal E Plane...
Page 34: ...Appendix A Approved Antennas Page 26 Apprion Proprietary Confidential E Plane...
Page 38: ......