3-4 Instrument Setup
Fuse Service
Configuration
The instrument is shipped with single-line service configuration. If your installation
requires double-line service, you must remove the jumper clip. See “Converting from a
Single to Double-Line Fuse” on page 3-5.
ELECTRICAL HAZARD. To protect yourself against shock hazards, use a
properly wired three-terminal outlet. Do not use an adapter to a two-terminal outlet.
ELECTRICAL FIRE HAZARD. Improper fuses can damage the wiring system
and cause a fire. Before turning on the instrument, verify that the fuses are properly installed.
Input Voltage
IMPORTANT
You must be able to disconnect the main power supply to the instrument
immediately if necessary.
The following table specifies the electrical operating range for the instrument in
various parts of the world. Select the appropriate fuse configuration based on the
voltage used.
Note
The Volt-Amp number for this instrument is 725 Volt Amps.
Interchangeable
Sample Block
Modules
The instrument features an interchangeable sample block module allowing portability
and versatility in sample configuration.
Note
The GeneAmp PCR System 9700 will not operate without an interchangeable sample
block module installed.
For an example, refer to the GeneAmp PCR System 9700 96-Well Sample Block
Module User Guide, for instructions on how to install a block and configure the
samples.
WARNING
!
WARNING
!
Location
Voltage (VAC)
a
a. Acceptable AC line voltage tolerances: 100, 120, 220, 230 ±10%; 240 VAC +6%/–10%, 50/60 Hz ± 1%.
Frequency
Amperage (A) Nominal
Japan
100 ±10%
50/60 Hz ±1%
3.16
USA/Canada
120 ±10%
50/60 Hz ±1%
4.20
EC
230 ±10%
50/60 Hz ±1%
3.14
Sample Block Module
Part Number
96-Well Gold Sample Block Module
4314443
96-Well Aluminum Sample Block Module
4314445
60-Well 0.5 mL Sample Block Module
4309131
Dual 384-Well Sample Block Module
N805-0400
Auto-Lid Sample Block Module
4312904
Summary of Contents for GeneAmp PCR System 9700
Page 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Page 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Page 10: ......
Page 28: ......
Page 56: ......
Page 90: ......
Page 96: ......
Page 110: ......
Page 116: ......
Page 124: ......
Page 126: ......
Page 131: ......