Manual Calibration
Data Explorer
™
Software User’s Guide
5-19
5
The Edit/Create Reference Peak Information dialog box
(Figure 5-6) is displayed.
Figure 5-6 Edit/Create Reference Peak Information
Dialog Box
2.
Type the Name and Theoretical m/z for a reference
compound, then select the charge state. Optionally, enter
the Elemental Composition for the compound.
3.
Specify the mass type (Resolved Isotope Mass or
Average Mass).
If you are adding masses for highly charged
compounds, see “Specifying mass type for highly
charged, narrow peaks” on page 5-21.
4.
Click Insert.
5.
Repeat step 2 through step 4 for remaining compounds.
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...