Chapter 3 Peak Detection and Labeling
3-22
Applied Biosystems
3
Basic Settings
(spectrum data
only)
Table 3-2 describes the parameters in the Basic Settings tab
of the Spectrum Peak Detection Setup dialog box (see
Figure 3-4 on page 3-14).
Default peak detection values are listed in Section 3.7, Default
Peak Detection Settings.
Table 3-2 Basic Settings Tab Parameters (Spectrum Data Only)
Parameter
Description
Global Thresholds
%Base Peak
Intensity
NOTE: This
parameter was
previously named
Peak Threshold%
and Base Peak
Relative%.
Specifies a percentage of the base peak intensity as the
threshold value.
To be detected, peaks must be above this threshold and
above the %Max Peak Area value.
If you override Global Thresholds by selecting Use Advanced
Settings (described on page 3-23), the software ignores
Global Thresholds and uses the %Base Peak Intensity,
%Max Peak Area, Minimum Area, and Minimum Intensity
thresholds set on the Advanced tab (described on page 3-28)
to detect peaks for the selected detection range.
NOTE: If you are examining Voyager data that contains a
Low Mass Gate spike, the software may identify the Low
Mass Gate spike as the Base Peak. Truncate the data as
described in Section 5.9, Truncating a Spectrum, to eliminate
the Low Mass Gate spike and correctly identify the Base
Peak.
Hint: For most Voyager applications, leave this parameter
set at 0 and adjust the %Max Peak Area.
Continued
Summary of Contents for Data Explorer 4 Series
Page 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Page 10: ...Table of Contents x Applied Biosystems ...
Page 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Page 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Page 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Page 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Page 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Page 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Page 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Page 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Page 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...