Handler
29
12 COMP.3
13 COMP.4
14 DISCH
Discharge.
(Low)
15
CHARG
Charge and Test. (Low)
17 TRIG8V
External Trigger Signal
(Rising edge)
17-18 SHORT = TRIG5V
18
TRIG24V
Table 5-2
Comparator Record Selection Truth Table
COMP
4-0
Record
COMP
4-0
Record
COMP
4-0
Record
COMP
4-0
Record
11111 No
change
10111
8 01111 16 00111 24
11110 1 10110
9 01110
17 00110 25
11101 2 10101
10 01101
18 00101 26
11100 3 10100
11 01100
19 00100 27
11011 4 10011
12 01011
20 00011 28
11010 5 10010
13 01010
21 00010 29
11001 6 10001
14 01001
22 00001 30
11000 7 10000
15 01000
23 00000 No
change
Table 5-3
Output Signals
4 EX0
See
Tale
5-4.
5 EX1
6 EX2
7 EX3
7
EOC
Measurement completion signal. (Low)
19
GD
Pass signal. (Low)
20
NG
Fail signal. (Low)
Table 5-4
Typical Voltage Control Signal
EX3 EX2 EX1 EX0
Voltage
1 1 1 0
50V
1 1 0 1
100V
1 0 1 1
250V
0 1 1 1
500V
Table 5-5
Power Signal
27
IN-GND
Internal GND: Not Recommend to use
28
IN-GND
29 EX-GND
External GND
30 EX-GND
32 EX-VCC2
External VCC1: Pull-up Resistance(5k
) Power Supply
33
EX-VCC1
External VCC2: Main Power Supply