[SETUP] Key
21
5.3.1
Comparator Nominal [R-NOMINAL],[V-NOMINAL]
SCPI Command: COMParator:TOLerence:RNOMinal <
float
>
SCPI Command: COMParator:TOLerence:VNOMinal <
float
>
When using the ABS and PER mode, the nominal value must be inputted.
The field is ignored when in SEQ mode.
The Nominal value must be positive.
◼
Procedure of inputting nominal value
:
Step 1
Press [Setup] key to enter <SETUP> page
Step 2
Use cursor key to select [R-NOMINAL] or [V-NOMINAL] field.
Step 3
Input the nominal value by numeric keypad.
5.3.2
Limit [R-LOWER] [R-UPPER] [V-LOWER], [V-UPPER]
SCPI Command: COMParator:TOLerence:RLMT
<lower>,<upper limit
>
SCPI Command: COMParator:TOLerence:VLMT
<lower>,<upper limit
>
The high limited value should be greater than low nominal value.
When in SEQ Mode, please input the direct value of resistance or voltage.
When in ABS
Mode, please input the absolute value of resistance or voltage.
When in PER
% Mode, please input the relative value of resistance or voltage
in %.
◼
Procedure of inputting high/low limited values:
Step 1
Press [Setup] key to enter < SETUP> page
Step 2
Use cursor keys to select :
[R-LOWER] [R-UPPER] or
[V-LOWER] [V-UPPER] field
Step 3
Please input the percentage value in when [PER
%] mode
Please input the Direct value when in ABS
and SEQ mode.