AT5220 User
’
s Guide
10.
Specifications
This chapter describes:
Technical Specifications
General Specifications
Dimensions
10.1
T
echnical Specifications
The data as below is acquired by the following conditions:
Temperature
:
23
°
C
±
5
°
C
Humidity
:
≤
65% R.H.
Adjustment
:
short-circuit clear zero before measurement
Warm-up Time
:
>60 minutes
Calibration validity
:
12 months
Channel
:
CH1
Measuring speed
:
slow
RANGE
Max Reading Resolution FAST
MED
SLOW
1
300m
Ω
300.00m
Ω
10
µ
Ω
1%
±
5 dgt
0.5%
±
3dgt 0.5%
±
2dgt
2
3
Ω
3.0000
Ω
100
µ
Ω
1%
±
5 dgt
0.5%
±
3dgt 0.5%
±
2dgt
3
30
Ω
30.000
Ω
1m
Ω
1%
±
5 dgt
0.5%
±
3dgt 0.5%
±
2dgt
4
300
Ω
300.00
Ω
10m
Ω
1%
±
5 dgt
0.5%
±
3dgt 0.5%
±
2dgt
5
3k
Ω
3.0000k
Ω
100m
Ω
1%
±
5 dgt
0.5%
±
3dgt 0.5%
±
2dgt
10.2
G
eneral Specifications
Display
:
TFT-LCD true-color display
,
3.5 inches
Measuring speed
:
All channels are turned on, range mode is manual:
SLOW
:
9s/20 channels
MED
:
6s/20 channels
FAST
:
4s/20 channels
Max reading
:
30000
Range
:
Manual
Calibration
:
short-circuit clear zero for all ranges
File
:
parameters are AUTO saved
Trigger
:
Internal
、
External
、
Manual and BUS
Interface
:
PLC
(
Handler
)
interface
RS232 interface
Programming language
:
SCPI
Auxiliary function
:
keypad lock