![Applent Instruments AT2811 User Manual Download Page 14](http://html.mh-extra.com/html/applent-instruments/at2811/at2811_user-manual_2954872014.webp)
AT2811 User
’
s Guide
14
All menu option is as below:
Setup
STD
Principal parameter nominal value
SEC
AUX parameter (if principal parameter is L,R, or if Z display q-L;if principal
parameter is C display d-H)
P1-H
Bin 1 upper limit (input positive percentage, such as: 5% , principal
parameter window displays 5.000)
P1-L
Bin 1 lower limit (input negative percentage, such as: -5%,principal
parameter window displays -5.000)
P2-H
Bin 2 upper limit (input positive percentage, such as: 10%,principal
parameter window displays 10.000)
P2-L
Bin 2 lower limit (input negative percentage,such as:-10%,principal
parameter window displays -10.00)
P3-H
Bin 3 upper limit (input positive percentage, such as:15%,principal
parameter window displays 15.000)
P3-L
Bin 3 lower limit (input positive percentage, such as:-15%,principal
parameter window displays -15.00)
BEEP
Beep setup:
OFF:
turn off
P1:
PASS bin P1 beep
P2:
PASS bin P2 beep
P3:
PASS bin P3 beep
AU:
AUX parameter FAIL beep
NG:
Principal and AUX parameter is all FAIL beep
SrES
Source Resistance setup:
100:
100
Ω
Source Resistance
30:
30
Ω
Source Resistance
LEVEL
Signal Level setup:
0.1:
0.1V
0.3:
0.3V
1.0:
1.0V
EQU
Equivalent Circuit setup:
SER:
series
equivalent
PAL:
parallel
equivalent
CAL
System correction page(password protection,is not open to users)
3.11.1
To setup sorting
Before set up sorting, please understand sorting system of AT2811:
AT2811 has enhanced sorting distinguish system.
P1,P2,P3 is used for indicating principal parameter PASS or not, if FAIL, NG indicator is lighted,
sorting finishes. If PASS, GD indicator P1-P3 is lighted, and continue performing AUX
parameter compare.
AUX is used for indicating AUX parameter PASS or not, if FAIL, AUX indicator is lighted, NG
indicator is extinguished.
If there is one NG in principal and AUX parameter, NG indicator is lighted.