![APPA A1 User Manual Download Page 9](http://html.mh-extra.com/html/appa/a1/a1_user-manual_2954775009.webp)
A1
EN
1-2 Environmental Conditions
Indoor Use.
Maximum operating altitude :
2000m
(6562ft)
Operating temperature :
0°C ~ 30°C, ≤ 80%RH
30°C ~ 40°C, ≤ 75%RH
40°C ~ 50°C, ≤ 45%RH
Storage temperature :
-20 to +60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, >
28°C
Pollution Degree :
2
Shock vibration :
MIL-PRF-28800F for A class
2 Instrument
Drop Protection :
4 Feet Drop to hardwood on concrete Floor
1-3 Electrical Specifications
Accuracy is given as ±(% of r counts
of least significant digit) at 23°C ± 5°C,with
relative humidity Less than 80% R.H. and is
specified for one year after calibration.
ACV and ACA specifications are ac
coupled,true R.M.S. The crest factor may be
up to 3.0 as 4000 counts.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0.
Add 5.0% for C.F. 2.0 ~ 2.5.
Add 7.0% for C.F. 2.5 ~ 3.0.
Position Error of Clamp: ±1.5% of LCD
reading.
8
Summary of Contents for A1
Page 4: ...A1 EN Power on off Auto Power Off Auto Power Off Disable Hold down and press 3 ...
Page 17: ...A1 TC 開啟 關閉背光功能 更換電池 Voltsense 手握量表 時 請勿超過 擋板 護手擋板 LCD 顯示的斜槓數量代表電場的強度 若未顯示 也可能存在電壓 按住 16 ...
Page 26: ...A1 SC 開啟 關閉背光功能 更換電池 Voltsense 手握量表 時 請勿超過 擋板 護手擋板 LCD 顯示的斜槓數量代表電場的強度 若未顯示 也可能存在電壓 按住 25 ...