9
Using the Store function
Store/Recall button
:
The screen will be flashing MEM symbol and the number of
stored data twice when the Store/Recall button is pressed.
1. Press once to store the voltage value during voltage test
2. Press once to store the PI/ DAR, insulation or Earth-bond
resistance value when the individual test is completed.
3. The memory is divided into five segments
,
each segments
has 100 maximum data. The method to store the data is first
in /first out when the memory is full.
Table: The stored values of the individual test
segments
Items
Voltage
Earth-Bond
resistance
Insulation
resistance
50~1000V
DAR
PI
1
Voltage Resistance Resistance
DAR
value
PI value
2
Leakage
current
R30-sec R1-min
3
Test voltage R1-min R10-min
Search the stored value under RECALL mode
NOTE :
Press the Store/Recall button
≧
5 sec to clear the stored data.
The screen will be flashing MEM symbol and “
cLr
” on
reading twice.
Using the Recall function
Press the Store/Recall button
≧
2 sec to enter/exit the RECALL
mode.
If the memory is empty, the meter will display the “nOnE ”
symbol.