15
(7) Maximum Hold
7-1
In maximum hold function the accuracy is changed as following.
Original Ac 10 digitals/ change steps of range .
For example :
At First , the maximum hold reading on display is 100.0mV on 400.0mV range. If a voltage vibration changes the
maximum hold reading to 120.0V. The change steps of range are 3 steps ( 400.0mV to 4.000V to 40.00V to
400.0V) so the accuracy is needed to add 3 change steps of range x 10 digitals / change steps of range = 30 digits.
7-2
On maximum hold mode the accuracy of resistance is specified from 400.0
Ω
to 400.0k
Ω
range only.
(8) Auto Power Off
The meter will automatically shut itself off after approximately 30 minutes after power on.
(9) Auto Power Off Disable :
Pressing and hold the ZERO A
Ω
<
button then turn the rotary switch from power off to
Ω
/
<
position quickly
(less than one second) that will disable the auto power off.