8
TROUBLESHOOTING YOUR SYSTEM
1.
Turbo unit turned ON. Air blows out of T6000 Spray Gun?
This is correct. Air blows constantly when Turbo unit is ON.
2.
Turbo unit turned ON. Unit does not operate or blow air?
Check that the unit is plugged in to proper electric source. Be sure there is power at the
electric source.
If there is no problem with electric supply, call for technical support.
3.
Turbo unit is turned ON, the trigger is pulled back on Spray Gun, but no solution comes out?
Check that the cup is screwed firmly to Spray Gun.
Check that there is no solution bubbling or leaking around the cup.
Turn Turbo unit OFF. Unscrew cup. Be sure that Solution Tube and Disc (#14) are
connected and have not come loose. Locate the two small holes in the Disc around
the Solution tube. Be sure that these two small holes are not clogged with dried
solution. To clean, carefully insert a pin through each hole ensuring that they are clear.
Turn Turbo unit OFF. Remove air cap ring, remove air cap, remove Air Distributor Plate.
Remove Solution nozzle (#3) with an adjustable wrench or spanner. There is a Solution
nozzle gasket (#4) on the Solution nozzle. Do not lose it. Rinse Solution Nozzle under
running water and check for dried solution that may have blocked the hole. Reinstall
Solution Nozzle Gasket. Reinstall Solution Nozzle, Air distributor plate, air cap and air
cap ring. Try spraying again. If you still have a problem call for technical support.
4.
Solution leaks in front of trigger.
With an adjustable wrench or spanner tighten Solution Needle Packing Screw (#7). Do
not over tighten. This can cause Solution needle to stick and not spring back. Try
spraying again. If leaking continues the Solution Needle Packing needs to be replaced.
(Rare). Remove Solution Needle Packing Screw (#7). Pry out the Solution Needle
Packing’s (4). Replace. Reinstall Solution Needle Packing Screw. Tighten
all the way, and then back off slightly. (1/8 turn) Test spray gun for leakage. If you still
have a problem call for technical support.
5.
When Turbo unit is ON, Solution continues to spray from Spray Gun without trigger being
pulled.
Immediately turn Turbo unit OFF.
Disconnect spray gun from air hose.
Slightly adjust (loosen) Solution Needle Packing Screw (#7). Test to see if problem is
resolved.
Check to see that Needle Spring is installed. (#9). Remove Solution Flow Screw (#10).
Look for Needle Spring around the back of the Solution Needle. (See diagram). If lost,
call for replacement.
Check Solution nozzle. Remove and clean following directions 9.c.iv. above.
6.
Solution leaks around the top of the cup.
Make sure the cup is screwed firmly to the body of the Spray Gun.
Make sure the threads around the cup are clean and the threads inside the cup top
assembly are clean.
Replace the cup top gasket. Pry out old gasket. Install new gasket. Test spray gun.
Summary of Contents for Mighty-Mist T6000
Page 9: ...9...