![Apollo DiaCheck DA12 User Manual Download Page 33](http://html1.mh-extra.com/html/apollo/diacheck-da12/diacheck-da12_user-manual_2954292033.webp)
31
DiaCheck DA12
BLOOD GLUCOSE MONITORING SYSTEM
Message
Symbo
l
Meaning
ean
Action
Temperature
out of range
You may have tested
near the low or high
end of the meter’s
operating temperature
range.
(10~40 degree Celsius
/50~104 degrees
Fahrenheit)
Repeat the test after the meter
has reached a temperature
within the operating range.
Low Battery
The power is too low to
perform a test.
Replace the batteries.
Lack of blood
volume
Applied blood sample
was too small.
Replace another new strip,
and apply enough blood
sample.
(>0.6μl)
Strip Damage
The strip could be
broken and cannot be
used
Please test with a new strip