MAINTENANCE AND RECALIBRATION 

Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated 
materials on the diffuser of the upward-looking sensor can block the optical path in three common ways: 

1.

 

Moisture or debris on the diffuser. 

2.

 

Dust during periods of low rainfall. 

3.

 

Salt deposit accumulation from evaporation of sea spray or sprinkler irrigation water. 

Apogee Instruments upward-looking sensors have a domed diffuser and housing for improved self-cleaning from 
rainfall, but active cleaning may be necessary. Dust or organic deposits are best removed using water, or window 
cleaner, and a soft cloth or cotton swab. Salt deposits should be dissolved with vinegar and removed with a cloth 
or cotton swab. 

Salt deposits cannot be removed with solvents such as alcohol or acetone.

 Use only gentle 

pressure when cleaning the diffuser with a cotton swab or soft cloth to avoid scratching the outer surface. The 
solvent should be allowed to do the cleaning, not mechanical force. 

Never use abrasive material or cleaner on the 

diffuser. 

It is recommended that sensors be recalibrated every two years. See the Apogee webpage for details regarding 
return of sensors for recalibration (

http://www.apogeeinstruments.com/tech-support-recalibration-repairs/

). 

 

Summary of Contents for SQ-627

Page 1: ...INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL EXTENDED RANGE PFD SENSOR Model SQ 627 in...

Page 2: ...of Compliance 3 Introduction 4 Sensor Models 5 Specifications 5 Deployment and Installation 8 Cable Connectors 8 Operation and Measurement 10 Maintenance and Recalibration 14 Troubleshooting and Cust...

Page 3: ...of electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the produc...

Page 4: ...quantum sensors but measure a wider range of wavelengths can be thought of as an extended range quantum sensor Typical applications of traditional quantum sensors include incoming PPFD measurement ov...

Page 5: ...respective manuals Model Signal SQ 620 Self powered SQ 622 0 2 5 V SQ 624 4 20 mA SQ 625 0 5 V SQ 626 USB SQ 627 SDI 12 SQ 628 Modbus Sensor model number and serial number are located on the bottom o...

Page 6: ...onse Time 0 6 s time for detector signal to reach 95 following a step change fastest data transmission rate for SDI 12 circuitry is 1 s Field of View 180 Spectral Range 340 to 1040 nm 5 nm wavelengths...

Page 7: ...respectively Mean spectral response measurements of six replicate Apogee SQ 600 series Extended Range PFD Sensors Spectral response measurements were made at 10 nm increments across a wavelength range...

Page 8: ...nted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere Azimuth error is typically less than 0 5 but it is easy to minimize by proper cable o...

Page 9: ...please contact Apogee directly to ensure ordering the proper pigtail configuration Alignment When reconnecting a sensor arrows on the connector jacket and an aligning notch ensure proper orientation D...

Page 10: ...tion php version 1 4 August 10 2016 Overview During normal communication the data recorder sends a packet of data to the sensor that consists of an address and a command Then the sensor sends a respon...

Page 11: ...ommand signals a measurement sequence to be performed Data values generated in response to this command are stored in the sensor s buffer for subsequent collection using D commands Data will be retain...

Page 12: ...ees if pointed down Available in sensors with serial number 3033 or greater aCC or aCC0 a00101 cr lf Returns mol m 2 s 1 w CRC aCC1 a00101 cr lf Returns millivolt output w CRC aCC2 a00101 cr lf Return...

Page 13: ...g the sensor version number and xx xx is serial number Running Average Command The running average command can be used to set or query the number of measurements that are averaged together before retu...

Page 14: ...g may be necessary Dust or organic deposits are best removed using water or window cleaner and a soft cloth or cotton swab Salt deposits should be dissolved with vinegar and removed with a cloth or co...

Page 15: ...ble Measurement Devices Dataloggers Controllers Meters Any datalogger or meter with SDI 12 functionality that includes the M or C command An example datalogger program for Campbell Scientific datalogg...

Page 16: ...warranty items to our factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its...

Page 17: ...eeinstruments com to discuss repair or replacement options OTHER TERMS The available remedy of defects under this warranty is for the repair or replacement of the original product and Apogee Instrumen...

Reviews: