Apogee SQ-600 Series Owner'S Manual Download Page 4

INTRODUCTION 

Radiation that drives photosynthesis is called photosynthetically active radiation (PAR) and is typically defined as 
total radiation across a range of 400 to 700 nm. PAR is almost universally quantified as photosynthetic photon flux 
density (PPFD) in units of micromoles per square meter per second (µmol m

-2

 s

-1

, equal to microEinsteins per 

square meter per second) summed from 400 to 700 nm (total number of photons from 400 to 700 nm). However, 
ultraviolet and far red photons outside the defined PAR range of 400-700 nm can also contribute to photosynthesis 
and influence plant responses (e.g., flowering). 

Sensors that measure PPFD are often called quantum sensors due to the quantized nature of radiation. A quantum 
refers to the minimum quantity of radiation, one photon, involved in physical interactions (e.g., absorption by 
photosynthetic pigments). In other words, one photon is a single quantum of radiation. Sensors that function like 

traditional quantum sensors, but measure a wider range of wavelengths can be thought of as an ‘extended range’ 

quantum sensor. 

Typical applications of traditional quantum sensors include incoming PPFD measurement over plant canopies in 
outdoor environments or in greenhouses and growth chambers, and reflected or under-canopy (transmitted) PPFD 
measurement in the same environments. The Extended Range PFD Sensor detailed in this manual uses a detector 
that is sensitive to radiation up to about 1100 nm, well beyond the range of wavelengths that influence 
photosynthesis and plant responses. 

This means this particular sensor should only be used for photon flux 

density measurements under LEDs.

 

Apogee Instruments SQ-600 series Quantum Light Pollution Sensors consist of a cast acrylic diffuser (filter), 
photodiode, and signal processing circuitry mounted in an anodized aluminum housing, and a cable to connect the 
sensor to a measurement device. SQ-600 series sensors are designed for continuous photon flux density 
measurements in indoor environments under LEDs. The SQ-640 Quantum Light Pollution models output a voltage 
that is directly proportional to photon flux density. The voltage output by the sensor is directly proportional to 
radiation incident on a planar surface (does not have to be horizontal), where the radiation emanates from all 
angles of a hemisphere.

 

Summary of Contents for SQ-600 Series

Page 1: ...NC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL QUANTUM LIGHT POLLUTION SENSOR Models SQ 640 including SS models ...

Page 2: ...te of Compliance 3 Introduction 4 Sensor Models 5 Specifications 5 Deployment and Installation 8 Cable Connectors 8 Operation and Measurement 9 Maintenance and Recalibration 13 Troubleshooting and Customer Support 14 Return and Warranty Policy 15 ...

Page 3: ...nt of electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products manufactured by us do not contain as intentional additives any of the restricted materials including lead see note below mercury cadmium hexavalent chromium polybrominated biphenyls PBB polybrom...

Page 4: ...uantum sensor Typical applications of traditional quantum sensors include incoming PPFD measurement over plant canopies in outdoor environments or in greenhouses and growth chambers and reflected or under canopy transmitted PPFD measurement in the same environments The Extended Range PFD Sensor detailed in this manual uses a detector that is sensitive to radiation up to about 1100 nm well beyond t...

Page 5: ...ir respective manuals Model Signal SQ 640 Self powered SQ 642 0 2 5 V SQ 644 4 20 mA SQ 645 0 5 V SQ 646 USB SQ 647 SDI 12 SQ 648 Modbus Sensor model number and serial number are located on the bottom of the sensor If the manufacturing date of a specific sensor is required please contact Apogee Instruments with the serial number of the sensor ...

Page 6: ...stability Less than 2 per year Non linearity Less than 1 up to 4000 µmol m 2 s 1 Response Time Less than 1 ms Field of View 180 Spectral Range 340 to 1040 nm 5 nm wavelengths where response is greater than 50 see Spectral Response below Directional Cosine Response 2 at 45 zenith angle 5 at 75 zenith angle see Directional Response below Azimuth Error Less than 0 5 Tilt Error Less than 0 5 Temperatu...

Page 7: ...ely Mean spectral response measurements of six replicate Apogee SQ 600 series Quantum Light Pollution Sensors Spectral response measurements were made at 10 nm increments across a wavelength range of 300 to 1100 nm in a monochromator with an attached electric light source Measured spectral data from each quantum light pollution sensor were normalized by the measured spectral response of the monoch...

Page 8: ...unted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere Azimuth error is typically less than 0 5 but it is easy to minimize by proper cable orientation In addition to orienting the cable to point toward the nearest pole the sensor should also be mounted such that obstructions e g weather station tripod tower or other instrumentation do no...

Page 9: ...ease contact Apogee directly to ensure ordering the proper pigtail configuration Alignment When reconnecting a sensor arrows on the connector jacket and an aligning notch ensure proper orientation Disconnection for extended periods When disconnecting the sensor for an extended period of time from a station protect the remaining half of the connector still on the station from water and dirt with el...

Page 10: ...f approximately 0 200 mV is required to cover the photon flux density range the sensor is capable of measuring In order to maximize the measurement resolution and signal to noise ratio the signal input range of the measurement device should closely match the output range of the PFD sensor Wiring for SQ 640 White Positive signal from sensor Black Negative signal from sensor Clear Shield Ground ...

Page 11: ... 1 µmol m 2 s 1 per mV Sensor Output Signal mV Photon Flux Density µmol m 2 s 1 1 50 50 Example of photon flux density measurement with an Apogee model SQ 640 Quantum Light Pollution Sensor The LED in this example Full yields a photon flux density on a horizontal plane of 50 µmol m 2 s 1 This yields an output signal of 50 mV This signal is converted to photon flux density by multiplying by the cal...

Page 12: ... SQ 620 Extended Range Quantum Sensor can measure a photon flux density range of 0 4000 mol m 2 s 1 The model SQ 640 Quantum Light Pollution Sensor is designed for low light measurements and can only measure a photon flux density range of 0 200 mol m 2 s 1 the sensor will output values beyond 200 mol m 2 s 1 but these are beyond the linear range of the sensor and is referred to as the low light ve...

Page 13: ...ation water Apogee Instruments upward looking sensors have a domed diffuser and housing for improved self cleaning from rainfall but active cleaning may be necessary Dust or organic deposits are best removed using water or window cleaner and a soft cloth or cotton swab Salt deposits should be dissolved with vinegar and removed with a cloth or cotton swab Salt deposits cannot be removed with solven...

Page 14: ...ng a sensitivity of 1 mV per µmol m 2 s 1 Thus a compatible measurement device e g datalogger or controller should have resolution of at least 1 mV in order to provide photon flux density resolution of 1 µmol m 2 s 1 and resolution of at least 0 1 mV in order provide photon flux density resolution of 0 1 µmol m 2 s 1 An example datalogger program for Campbell Scientific dataloggers can be found on...

Page 15: ...d warranty items to our factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating range 3 Natural occurrences such as lightning fire etc 4 Unauthorized modification 5 Improper or unauthorized repair Please note that nominal accuracy drift is normal over time R...

Page 16: ...geeinstruments com to discuss repair or replacement options OTHER TERMS The available remedy of defects under this warranty is for the repair or replacement of the original product and Apogee Instruments is not responsible for any direct indirect incidental or consequential damages including but not limited to loss of income loss of revenue loss of profit loss of wages loss of time loss of sales a...

Reviews: