background image

5

8

6

7

(SMPL)

99 Sample

Measurements

(LOG)

99 Log

Measurements

(LOG)

99 Daily Averages

Measurements

Viewable on meter LCD & downloadable

Downloadable Only

Cleaning

Debris on the meter is a common cause of low readings.  Salt 

deposits can accumulate on a sensor from evaporation of 

sprinkler irrigation water and dust, which can accumulate 

during periods of low rainfall. Salt deposits should be dissolved 

and removed with vinegar and a soft cloth or q-tip. Dust and 

other organic deposits are best removed with water, rubbing 

alcohol or window cleaner. 

Temperature Sensitivity

A change in temperature changes the amount of O

2

 available 

to the sensor and therefore changes the mV output that 

correlates to the atmospheric constant of 20.95%. Additionally, 

the sensor electronics have a small temperature dependence. 

To eliminate temperature effects, simply recalibrate.  

Pressure Sensitivity

A change in barometric pressure changes the amount of O

2

 

available to the sensor and therefore changes the mV output 

that correlates to the atmospheric constant of 20.95%. To 

eliminate pressure effects, simply recalibrate.

Using the Meter

1) Press the power button to start. The meter 

will turn itself off two minutes after button is 

pushed to conserve battery.

2) To calibrate sensor: While sensor is in 

open air, press 

mode

 twice (RUN should be 

blinking), then while pressing 

sample

, press 

mode

 once. Screen will display CAL during 

calibration. Calibration can be done in either 

SMPL or LOG mode. Recalibration should be done 

when significant changes in pressure, humidity, 

or temperature occur. See page 3 for further 

information about these effects on output. See 

page 10 for life expectancy.

3) To choose mode: To choose 

between SMPL and LOG modes 

push 

mode

 once and use the 

up 

and

 down

 

arrows to make choice. Once desired mode is 

blinking press 

mode

 two more times to begin.

4) For automatic measurements: Use LOG mode 

(use step 3 to choose LOG mode). Meter will 

power on and off to record a measurement every 

thirty seconds.

NOTE:  O2 Meter units are % concentration.

5) For manual measurements: Use SMPL mode (use step 3 to 

choose SMPL mode). Press 

sample

 when you want to take a 

measurement. Store up to 99 manual measurements. When 

99 measurements are taken in SMPL mode, no more SMPL 

measurements can be stored.

6) To reset meter: From LOG or SMPL mode, press 

mode

 twice 

(RUN should be blinking), then while pressing down, press 

mode

 once. 

CAUTION, this will erase ALL measurements.

7) To review data: Press 

up/down

. To exit and return to 

present conditions and the capability to take measurements 

press sample.

8) In LOG mode, every 30 minutes it will average the 60 

30-second measurements taken and store the average. 99 

30-minute averaged measurements can be stored. Every 

48 measurements (making a 24 hour period) the meter will 

also store a daily average. In addition, 99 daily averages 

can be stored and are available for download only, these 

measurements are not viewable on the meter LCD.  All 

measurements taken in LOG mode will continue to be stored, 

eliminating the oldest measurement. To keep data, switch out 

of log mode.

Reviews: