EN.IX.APM.2
Revision 01.A.001/2022.09
9
8.1.
INTRINSICALLY-SAFE Exi TRANSMITTERS ACCORDING
TO CERTIFICATE JSHP 22 ATEX 0051X
8.2.
Standards used for assessment
The transmitters are manufactured in compliance with the following standards:
EN IEC 60079-0:2018; (IEC 60079-0:2017 ed. 7.0),
EN 60079-11:2012; (IEC 60079-11:2011 ed. 6.0.).
8.3.
Transmitter
explosion-proofing
designations
according
to
certificate JSHP 22 ATEX 0051X
The following ATEX markings apply only to intrinsically safe transmitters:
ATEX:
II 1/2G Ex ia
IIC T5/T4/T3 Ga/Gb
II 1D Ex ia IIIC T100
ºC/T135ºC/ T155ºC Da
JSHP 22 ATEX 0051X
As standard, the transmitters are manufactured for explosive gas atmospheres.
A customer can also order a transmitter for explosive dust atmospheres.
8.4.
Minimum supply voltage
Table 1.
Minimum supply voltage of an Exi transmitter.
Minimum supply voltage for
transmitters with backlight off
Minimum supply voltage for
transmitters with backlight on *)
13V
16V
*) Backlight can only be turned on at transmitter manufacturing
stage.
8.5.
Load resistance
a) For a linear power supply from a barrier:
𝑅
𝑜 max [Ω]
=
𝑈
𝑧𝑎𝑠
− 13𝑉 − (𝑅
𝑊
∗ 0.02066𝐴)
0.02066𝐴
b) For power supply from a trapezoidal or orthogonal source.
𝑅
𝑜 max [Ω]
=
𝑈
𝑧𝑎𝑠
− 13𝑉
0.02066𝐴
Rw
– baririer resistance;
Uzas
– minimum voltage of the barrier used.