Apera PC60-Z Instruction Manual Download Page 10

 

- 10 - 

 

Paramet er Setting  

 

Table of Settings 

 

Parameter Setting 

1) 

When the meter is turned off, long press           

to enter parameter setting →short press          to switch P01-P02… 

→P14. Short Press          , parameter flashes → short press            to adjust parameter → short press            to 

confirm →Short press          to exit parameter setting and go back to measurement mode. 

2) 

Auto. Lock (P02)

 

– Users can set the auto lock time from 5 to 20 seconds. For example, if 10 seconds is set, when 

the measured value is stable for more than 10 seconds, the measured value will be automatically locked, and the 

HOLD icon will be displayed. Short press          to release the lock. When the setting is "Off" , the Auto. lock function 

is turned off, that is, the measured value can only be manually locked. Short press          to lock or unlock the 

measured value. The HOLD icon will be displayed when reading is locked. 

3) 

Auto. Backlight (P03)

 

  Users can set the automatic backlight time for 1 to 8 minutes. For example, if 3 minutes 

is set, the backlight will turn off automatically after 3 minutes; when the 

“Off” is set, the auto. backlight function will 

be turned off, and short press          to manually turn the backlight on or off. 

4) 

Auto. Power off (P04)

 

  The auto. power off time can be set to 10 to 20 minutes. For example, if 15 minutes is 

set, the meter will automatically shut down after 15 minutes if no operation; when 

“Off” is set, the auto. power off 

function will be turned off. Long press          to manually shut down the meter. 

5) 

pH Calibration Reminder (P07) and Conductivity Calibration Reminder (P11) 

– set X hours (H) Or X days (D) 

in ZenTest mobile app 

– settings – Parameter – pH – Calibration Reminder. On the meter, you can only check the 

values that’s been set up on ZenTest App. For example, if 3 days is set up, the Er6 icon (see Figure-4) will appear 

in the lower right corner of the LCD screen in 3 days to remind you to perform calibration, also in the ZenTest App 

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C – ˚F 

˚F 

P2 

Select automatic lock 

5-20 seconds 

– Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

– Off 

P4 

Automatic Power Off 

10-20 minutes 

– Off 

10 

P5 

pH Buffer Series Selection 

USA 

– NIST 

USA 

P6 

pH Resolution 

0.1 

– 0.01 

0.01 

P7 

pH Calibration Reminder 

H-hours D-Days (set up in ZenTest App) 

P8 

pH back to factory default 

No

 

– Yes 

No 

P9 

Conductivity Reference Temperature 

1

5 ˚C to 30 ˚C 

2

5 ˚C 

P10 

Temp. Compensation Coefficient 

0 to 9.99 

2.00 

P11 

Conductivity Calibration Reminder 

H-hours D-Days (set up in ZenTest App) 

P12 

Conductivity Back to Factory Default 

No 

– Yes 

No 

P13 

TDS Factor 

0.40 to 1.00 

0.71 

P14 

Salinity Unit 

ppt 

– g/L 

ppt 

Summary of Contents for PC60-Z

Page 1: ... 1 SKU AI3719 PC60 Z Smart Multi Parameter Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de V1 03 ...

Page 2: ...d Functions 11 Probe Replacement 12 Warranty 12 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leaves factory It does NOT mean the product is used 2 The batteries are already preinstalled Just pull off the paper slip before using the tester When you replace the batteries make sure to foll...

Page 3: ...l display dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following graphic guides Self Diagnosis Er1 Er6 icons Detailed problem analysis and solutions Parameter Setup Press buttons to set up except for P7 and P11 All parameters can be set up in Settings Alarm T...

Page 4: ...urn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 In measurement mode when automatic lock is turned off short press to manually lock or unlock readings Pull the battery cap up S...

Page 5: ... will make the electrode response slow If this happens soak the pH electrode in 3M KCl solution for 3 5 hours and then re calibrate it before using 4 The storage solution is 3M KCL SKU AI1107 and the tester kit comes with a bottle of 10mL storage solution can be used repeatedly If it is contaminated replace it with a new one Please do not use other brands storage solutions as they may contain othe...

Page 6: ...use other buffer solutions to conduct 2nd or 3rd point calibration Perform the 2nd point calibration 4 00 pH immediately after the 1st point Do NOT turn off the meter before you perform 2nd point calibration If the meter is turned off after 1st point calibration users will need to restart the calibration process with the 7 00 pH or 6 86 pH first and the 2nd point following after Otherwise Er1 will...

Page 7: ...e please contact us at info aperainst de pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buffer solutions comply with the USA or NIST standard 2 Check whether pH buffers are expired or contaminated 3 Please check whether pH electrode is damaged or broken If so please replace with a new one 4 The electrode is aged used for over a year and has a much slower response...

Page 8: ...ication Icon Calibration Standards Measuring Range 84 μS cm 0 199 μS cm 1413 μS cm 200 1999 μS cm 12 88 mS cm 2 0 20 00 mS cm 3 The tester has been calibrated before leaving factory Generally users can use the tester directly or users can test conductivity calibration solutions first If the error is large then calibration is needed 4 Conductivity calibration solutions are easier to get polluted th...

Page 9: ...us at info aperainst de The calibration reminder is triggered It s time to perform a new conductivity calibration Perform conductivity calibration or cancel calibration reminder in ZenTest settings Conductivity Measurement Press key to turn on the tester Press to switch to Conductivity measuring mode Rinse the probe in distilled water and dry it Dip the probe in sample solution shake it for a few ...

Page 10: ...can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down after 15 minutes if no operation when Off is set the auto power off function will be turned off Long press to manually shut down the meter 5 pH Calibration Reminder P07 and Conductivity Calibration Reminder P11 set X hours H Or X days D in ZenTest mobile app settings Parameter pH Calibration Remi...

Page 11: ...n 0 50 C 32 122 F Conductivity Range 0 to 199 9 μS 200 to 1999 μS 2 to 20 00 mS cm Resolution 0 1 1 μS 0 01 mS cm Accuracy 1 F S Calibration Points 1 to 3 points TDS Range 0 1 ppm to 10 00 ppt TDS Factor 0 40 to 1 00 Salinity Range 0 to 10 00 ppt Resistivity Range 50Ω to 20MΩ ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Icons and Functions Ca...

Page 12: ...m defects in material and workmanship and agree to repair or replace free of charge at option of APERA INSTRUMENTS Europe GmbH any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal we...

Reviews: