DJS-0.1-C Conductivity Electrode Instruction Manual

 

 

Model

 

Cell 

constant

 

Electrode 

size

 

Dimension 

size

 

Socket

 

Electrode 

body

 

Cable 

length

 

Range

 

DJS-0.1-C

 

0.1

 

7×18mm 

platinum

 

Φ12×155

 

BNC

 

Glass

 

1m

 

(0-200) 

µ

S/cm

 

 

Use and maintenance:

 

1.1 

Connect the DHS-0.1-C conductivity electrode to your conductivity meter. Set the cell constant 

to 0.1. Then calibrate the meter in the lower calibration point (e.g. 84

 µ

S/cm).

 

1.2 

Keep the conductivity electrode clean, before and after the measurement, rinse the electrode 

with deionized water/distilled water. In order to maintain the highest measurement accuracy, we 

recommend rinsing the electrode several times with your test sample solution.

 

1.3 

A Flow cell (included) is recommended when testing ultra-low conductivity solutions like ultra-

pure water. It helps reduce the influence from contact with air and minimizes reading error. Use 

a rubber tube (7mm in diameter, not included) to connect your water source on the low end of 

the flow cell. When water flows through the glass cell, conductivity can be measured without air.

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
1.4 

How to clean a conductivity electrode:

 

(a) 

Use warm detergent water or alcohol to clean off organic contaminants on the electrode.

 

(b) 

Use 10% lemon acid to clean calcium and magnesium sediments.

 

1.5 

Conductivity electrode storage: Store dry when not in use.

 