- 8 - 

7.3

 How to take pH measurements 

Short press            to power on the tester. Rinse the probe in pure water, shake off excess water. Insert the probe in your 

sample solution or solid sample, and hold still. Record the reading when it is stabilized(          stays on screen). 

7.4 

Pure Water pH Measurement 

When testing pure water like tap water, drinking water, RO water and distilled water, it will take longer for the readings to 

get fully stabilized (typically 1-5 minutes). Please be patient. If still not working, add Apera 3M KCL (AI1107) to your pure 

water at the ratio of 1:1000 (e.g. 1 ml KCL to 1000 ml water) to accelerate stabilization while minimizing pH change. If the 

accuracy is not meeting your requirement, please contact us at [email protected] to find the specialized meter designed 

for pure water pH test. 

 

ORP Measurement  

1)  ORP stands for Oxidation-

Reduction Potential, measured in mV. It’s also called redox. ORP is a measure of the 

cleanliness of water & its ability to break down contaminants. A separate ORP probe (ORP60-DA) needs to be 

installed to be able to measure ORP.   

2)  Power on the tester, unscrew the original probe, and install the ORP60-DA probe, then the tester will automatically 

switch to ORP measurement mode (Refer to Section 15 for how to replace a probe). 

3)  Rinse the probe in distilled water and dry it. Dip the probe in sample solution, shake for a few seconds, and hold still. 

Record the ORP reading after            appears and stays on screen. 

 

 

Probe Cleaning  

1)  The tester is only as accurate as the probe is clean. Always thoroughly rinse off the probe before and after each 

measurement with pure water in a container or with a wash bottle.   

2)  For tough contaminants, detach the sensor shield, soak the probe in Apera probe cleaning solution (AI1166) or 

detergent water for 30 minutes. Then use a soft brush to remove the contaminants. Afterwards, soak the probe in 

3M KCL soaking solution for at least 1 hour. Rinse it off, then re-calibrate the tester before using again. 

 

Probe St orage  

1)  Always store the spear probe in the 3M KCL soaking solution  when not in use. Never let it dry out. 

Replace the soaking solution if it gets contaminated.  

2)  NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could damage the pH probe. 

If this happens, immediately soak the pH probe Apera 3M KCL soaking solution overnight, then re-calibrate it before 

using. Pure water is only for rinsing the probe. 

Summary of Contents for ZenTest PH60S-Z

Page 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Page 2: ...E 5 PH CALIBRATION 5 PH MEASUREMENT 7 ORP MEASUREMENT 8 PROBE CLEANING 8 PROBE STORAGE 8 PARAMETER SETTING 9 TECHNICAL SPECIFICATIONS 10 ICONS AND FUNCTIONS 10 TROUBLESHOOTING GUIDE 11 PROBE REPLACEME...

Page 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Page 4: ...to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press...

Page 5: ...water RO water or tap water is the alternative then shake off excess water see graph on the right 5 3 Perform at least a 2 point calibration For pH calibration tutorial refer to Section 6 5 4 When you...

Page 6: ...librate 2nd or 3rd point Otherwise you will need to restart the calibration process with 7 00 pH first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 1...

Page 7: ...d for over a year and has a much slower response A replacement is needed pH probe zero electric potential out of range 60mV or 60mV 1 Check whether pH buffer solutions comply with the USA or NIST stan...

Page 8: ...inal probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 15 for how to replace a probe 3 Rinse the probe in distilled water and dry...

Page 9: ...if 15 minutes is set the meter will automatically shut down after 15 minutes if no operation when Off is set the auto power off function will be turned off Long press to manually shut down the meter...

Page 10: ...unctions Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal B...

Page 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Page 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Reviews: