background image

 

- 12 - 

(approximately 18ml), see Diagram-4, tighten the vial cap. 

2)    Before measuring, gently shake the sample vial to make sample 

solution  even  and  wait  for  air  bubbles  to  disappear,  see 

Diagram-5, wait for 2~5 minutes until bubbles eliminate.   

3)    Wipe the sample vial surface with a microfiber cloth. Ensure that 

the vial is dry, clean and free from smudges. 

4)    Apply 1 drop of silicone oil on the surface of the vial. Wipe with 

a microfiber cloth to obtain an even distribution over the entire surface in order to eliminate scratches 

and smudges and improve light scattering, see Section 4.1(c) for details. 

 

5.3 

Measurement Modes 

a)  Normal Measurement Mode 

Press                button, the display shows progress bar (see diagram-6). It will display measured value 

after 8 seconds. Please press                  button again for repeated measurement. 

b)  Average Measurement Mode 

Long  press                button,  release  it  until  you  hear  a  beep,  then  the  instrument  enters  average 

reading  mode.  In  this  mode,  the  instrument  will  perform  10  times  of  continuous  measurement, 

displaying 1 data every 8 seconds, and will display the average value at the end, see diagram 6. The 

average measurement mode can be used for observing the stabilizing process of turbidity, and can 

also be used for testing rapid-settling solutions. 

 

 

 

 

 

 

 

5.4 

Notes for Measurement 

a) 

Keep the sample stable

: After the vial is placed into the sample cell, it is recommended to wait for 

1 to 2 minutes before calibration, as the solution will experience some shaking when the vial moves, 

which may affect measurement accuracy. 

b) 

Sample  vial  cleaning  requirement

:  Sample  vials  must  be  rigorously  cleaned  and  free  from 

smudges  or  scratches.  When  wiping,  user  should  grip  the  cap  and  bottom  to  avoid  leaving 

fingerprints on the surface of the vial. Its surface should be applied with a drop of silicone oil be wiped 

with a micro-fiber cloth. After that, please clean with  filter paper or high-quality tissue paper. See 

section 4.1(c) for details. 

c) 

Mixing and Degassing

: Samples should not be vigorously shaken or vibrated. It is recommended 

that users gently shake the sample vial to make solution evenly distributed. Air bubbles in solution 

Diagram- 4 

 

Diagram- 5 

 

Diagram 6 

 

Summary of Contents for TN400

Page 1: ...1 TN400 Portable Turbidity Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com...

Page 2: ...Display Mode 6 3 4 Keypad 7 3 5 Power Supply 7 4 Meter Calibration 8 4 1 Preparation for Calibration 8 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 9 4 3 Notes for Calibration 10 5 T...

Page 3: ...characteristics Smart functions such as automatic calibration operating instructions automatic power off low voltage display etc Large TFT color screen with blue background in measuring mode green bac...

Page 4: ...s AMCO Polymer or Formazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measurement Mode Normal measurement and average measurement mode Display TFT color screen Sample Vials 25...

Page 5: ...vial holder Close the cover when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark...

Page 6: ...0 NTU Carrying case TN400 Turbidimeter Microfiber cloth Screwdriver to remove battery cover Silicone oil 10 ml AA Alkaline battery 4 Sample vial 6 Instruction manual Diagram 2 Measurement mode Measure...

Page 7: ...erform single shot measurement long press the key to perform continuous measurement In calibration mode press the key to confirm the calibration In calibration mode press the key to select standard so...

Page 8: ...your supplier to replace See the detail in Section 7 Replacement Parts b Replace zero turbidity solution Replace with 0 0 NTU calibration solution open the vial cap pour out the original solution add...

Page 9: ...the vial to make the solution even For polymer calibration solutions that have not been used for a long time slowly flip the vial twice and let it stand for 2 minutes Be careful not to shake the solu...

Page 10: ...00 NTU calibration vial in the sample vial holder and press to calibrate After calibration is done press to confirm repeat the steps to calibrate 400 NTU and 800 NTU After calibration is done press to...

Page 11: ...ation solution If the wrong solution is selected for calibration the measurement will be completely wrong To fix this problem simply re calibrate the meter selecting the correct calibration solutions...

Page 12: ...s of continuous measurement displaying 1 data every 8 seconds and will display the average value at the end see diagram 6 The average measurement mode can be used for observing the stabilizing process...

Page 13: ...d for measurement Please do not wash the sample vial holder as it may damage the optical structure 6 WARRANTY We warrant this instrument to be free from defects in material and workmanship and agree t...

Page 14: ...L 1 bottle T200 1 20 100 400 800 NTU AMCO polymer Compatible with TN400 TN480 4 bottles T500 3 Sample vials with lid 25 60mm 1 6 6 pcs TN400 S3 Silicone oil 10mL 1 bottle APERA INSTRUMENTS LLC Address...

Reviews: