- 18 -
Table (6-1)
Measurement range
0.05 to 20 μS/cm
0.5 μS/cm to 200 mS/cm
Electrode constant
K = 0.1 (flow test)
K = 1.0
Calibration solution
84 μS/cm
84 μS/cm
1413 μS/cm
12.88 mS/cm
111.9 mS/cm
Calibration icon
Note: When you choose to use DJS-0.1-S pure water conductivity electrode (sold separately, K=0.1, no
temperature sensor built in), you can use Manual Temperature Compensation mode: long press ,
temperature value starts flickering, press or to change the value of temperature and then
press to confirm.)
7.6
Parameter Setting:
Conductivity measurement parameter setting Table:
Table (6-3)
Prompt Mark
Parameter Setting Items
Code
Parameter
P1
Standard solution series selection
USA (84μS/cm, 1413μS/cm,
12.88mS/cm, 111.9mS/cm)
CH (146.6μS/cm, 1408μS/cm,
12.85mS/cm, 111.3mS/cm)
P2
Electrode constant selection
0.1, 1, 10
P3
Reference temperature selection
25°C – 20°C – 18°C
P4
Temperature compensation
coefficient setting
0.00 to 9.99%
P5
Electrode constant setting
P6
Temperature unit setting
°C / °F
P7
Backlight display time setting
0-1-3-6min
P8
Auto power off time setting
0-10-20min
P9
Restore to producer setting
OFF-On