LabSen

®

 553 Professional Spear pH/Temp. Electrode 

User Manual 

 

Backed by proprietary Swiss sensor technology and materials, the LabSen 553 Professional Spear 
Combination pH Spear Electrode is designed for testing soil directly. 
 

 

Long-Life Reference System – composed of a glass tube, AgCl and a reference silver wire. The 
top end of the slim glass tube is stuffed with cotton, which prevents the reaction between AgCl 
and electrolyte when temperature changes. It improves the stability of reference electrode and 
prolong the service life of the electrode. 

 

Polymer Electrolyte – The solid polymer electrolyte makes the electrode hard to be 
contaminated and maintenance-free. 

 

Built-in Temperature Sensor 

 

Special Glass Membrane – The LabSen 553 pH electrode adopts a spear sensor tip, which is 
ideal for directly testing semi-solid/solid samples like soil. 

 

1. Technical Data 

Measuring Range

 

0-14

pH 

Electrolyte

 

3M KCl 

Temperature Range

 

0~60

 

Soaking Solution

 

3M KCl 

Shaft Material

 

PVC Plastic

 

Electrode Dimension

 

Electrode   

Φ15×90

mm 

Measuring Tip

Φ6×25

mm 

Membrane Shape

 

Conical 

Reference 

Long Life 

Connector 

BNC+RCA

 

Junction 

Single Pore 

Cable 

Φ3×1m 

 
 

2. Usage and Maintenance 

2.1

 

After  long  time  exposure  in  air,  Polymer  electrolyte  will  cause  shrinkage.  It  will  not  happen  in 
measurement status. The electrode must be soaked in 3M KCl solution for storage.   

2.2

 

The transparent polymer gel electrolyte will inflate and ooze out a bit when it’s soaked in KCL 
solution. This will not affect the measurement of the electrode, and it will happen more frequently 
with newer electrodes. This is actually a “self-clean” function especially when the junction gets 
contaminated. Just wipe off the inflated gel with clean tissue and continue to use the electrode as 
normal. 

2.3

 

The electrode’s measuring tip should be soaked in the soaking bottle containing storage solution to 
keep the membrane hydrated and junction unblocked. When measuring, please unscrew the bottle cap, 
pull out the electrode and rinse it with deionized or distilled water. After using, please put the electrode 
back into the bottle and screw tight the cap. Clean the bottle and replace the storage solution if the 
storage solution gets turbid and mildewed. The electrode should NEVER be soaked in purified water 
or buffer solutions for long. 

2.4

 

The connector of the electrode should  be kept clean and dry. If contaminated, please clean it with 
medical cotton and absolute alcohol and blow dry to prevent the short circuit of the electrode and slow 
reaction of electrode. 