20
7.4. Sub-menu of conductivity parameter setting (press key or key to switch)
P2.1
– Select electrode constant
(1.0-10.0-0.1)
1. In P2.0
mode, press key to enter in
P2.1 mode, please refer to the
left Diagram.
2. Press key,
1.0
blinks, then press key to select blinking 1.0
→
10.0
→0.1, when parameter blinks, press key to confirm.
3. After confirm the parameter, press key to enter in P2.2 mode, or
press key to return to the measurement mode.
P2.2
– Select conductivity standard solution
(Standard-CUS)
1. Press key,
Std
blinks, then press key select Std
→CUS, press
to confirm. Std
– Standard series, CUS – customer defined.
2. After confirm the parameter, press key to enter in P2.3 mode, or
press key to return to the measurement mode.
P2.3
– Select reference temperature
(15.0°C~30.0°C)
1. Press key,
25.0°C
blinks, then press or key to adjust
temperature value 15.0~30.0, press key to confirm.
2. After confirm parameter, press key to enter in P2.4 mode, or press
key to return to the measurement mode.
P2.4
– Adjust temperature compensation factor
(0.00 ~9.99%)
1. Press key,
2.00
blinks, press key or key to adjust
temperature compensation factor 0.00
–9.99, press key to confirm.
2. After confirm the parameter, press key to enter in P2.5 mode, or
press key to return to the measurement mode.
P2.5
– Adjust TDS coefficient
(0.40~1.00)
1.Press key,
0.71
blinks, press key or key to adjust TDS
Coefficient (0.40~1.00 ), press key to confirm.
2. After confirm the parameter, press key to enter in
P2.6 mode, or
Press key to return to the measurement mode.
P2.6
– Set due calibration
(No
– H00 – D00)
1. Press key,
no
blinks, then press key to select no
→H00→d00
Press to confirm.
2. When
H
blinks, press key,
00
blinks. Press key to adjust
Hours( 0~99 hours), press key to confirm; When
D
blinks. press
key,
00
blinks. Press key to adjust days (0~99 days), press
key to confirm.
3. After confirm parameter, press key to enter in P2.7 mode, or press
key to return to the measurement mode.