Apera Instruments DO8500 Instruction Manual Download Page 10

 

- 10 - 

(C)  Before the first use, unscrew the calibration sleeve to check if the sponge is damp. If 

the sponge is dry or if the electrode is exposed to dry air for more than 8 hours, the 

surface coating of the sensor cap may be dried out. So the electrode should be 

soaked in clean water at room temperature for 24 hours. If the water temperature is 

low, soaking time should be 48-72 hours. 

(D)  The sponge can not be allowed to get stained or moldy, otherwise it will consume or 

produce oxygen. If stained or moldy, please replace it immediately (some back-up 

sponges are included in the kit).   

4.3 

Sensor Cap 

(A)  The sensor cap is a key part of the optical DO probe. The surface coating of the cap 

cannot be scratched or mechanically worn. Otherwise, the service life of the sensor 

cap will be reduced or the probe will be damaged. Please pay special attention to it 

when using the probe. 

(B)  The surface coating of the sensor cap cannot withstand high temperature, so the 

optical DO probe can not be tested in water above 50

˚C/122˚F. 

(C)  If the surface of the sensor cap is contaminated, please do not use alcohol or organic 

solvents to clean, otherwise it may damage the probe. It can be gently wiped with a 

soft clean cloth. To disinfect the probe, immerse it in 3% hydrogen peroxide for 15 to 

30 minutes and then rinse off with clean water. 

(D)  The sensor cap has a service life of more than 8000 hours. When the probe is not 

being used, the service life will not be reduced, so the actual use time of the sensor 

cap is far more than a year. The major factor affecting the service life of the sensor 

cap is the surface coating being damaged under external force. So the key is to 

protect the sensor cap from external damage. 

(E)  If the sensor cap is damaged or deteriorated, users need to purchase a new one. 

Every new cap has a set of calibration codes which need to be input into the 

instrument. The specific input method will be described in the instruction manual of 

sensor cap. 

(F)  Users should not take off the sensor cap when it is not in use. Nor should one swap 

the caps from different instruments. When being installed, the sensor cap must be 

tightened, and the interior can not be contaminated or wet.   

 

Preparation for Calibration  

 

5.1 

Dissolved Oxygen Units Selection   

Dissolved oxygen unit displays in two formats: mg/L and %, and ppm and %. Press          to   

Summary of Contents for DO8500

Page 1: ...1 DO8500 Portable Optical Dissolved Oxygen Meter Instruction Manual APERA INSTRUMENTS LLC aperainst com v1 2...

Page 2: ...3 Sensor Cap 9 5 Preparation for Calibration 10 5 1 Dissolved Oxygen Units Selection 10 5 2 Resolution Selection 11 5 3 Temperature Unit Selection 11 5 4 Air Pressure Compensation 11 5 5 Salinity Comp...

Page 3: ...up to 8000 hours of service life Easy to calibrate and maintain Probe is equipped with a calibration storage sleeve which makes calibration and maintenance more convenient and reliable 1 2 Intelligent...

Page 4: ...tion Points Saturation Point Zero Oxygen Temperature Compensation Automatic 0 to 50 Air Pressure Compensation Automatic 60 to 120 kPa Salinity Compensation Manual or automatic 0 to 45 ppt Temperature...

Page 5: ...called Timing storage Temperature value and prompts of special display mode Date time and prompts of special display mode Automatic reading lock up Units of date and time ATC Auto temperature compensa...

Page 6: ...r 2 seconds to turn off Short Press In measurement mode press the key to switch unit mg L or ppm Long Press In the measurement mode press the key for 2 seconds to enter the calibration mode Short Pres...

Page 7: ...ll be displayed on the screen as shown in Figure 5 DO value 8 18mg L temp value 25 2 When salinity probe is connected and salinity is greater than 1ppt salinity will be displayed on the bottom right o...

Page 8: ...ashing and the meter stops the auto timing storage In auto timing mode manual storage does not work Set time to 00 00 in parameter P4 1 to exit from the auto timing storage mode 3 6 3 Recall stored va...

Page 9: ...hould be above this line 4 2 Probe Maintenance The sensor cap of the optical DO electrode must be kept in a moist environment If the surface coating of the sensor cap dries out the reading will be uns...

Page 10: ...ated please do not use alcohol or organic solvents to clean otherwise it may damage the probe It can be gently wiped with a soft clean cloth To disinfect the probe immerse it in 3 hydrogen peroxide fo...

Page 11: ...n general users do not have to calibrate air pressure any more If necessary calibrate it according to standard value measured by aneroid barometer Refer to parameter setting P3 5 for the procedure of...

Page 12: ...e refer to Figure 13 Press to return to measurement mode Special notes A Use 12 88mS cm conductivity calibration solution to calibrate salinity If the standard solution is wrong Err icon will be flash...

Page 13: ...ffective within 1 hour B Put the electrode into the oxygen free water wait for 3 to 5 minutes after the instrument is turned on and wait for the temperature and DO reading to completely stabilize The...

Page 14: ...influenced by temperature and the probe senses temperature slower than dissolved oxygen 8 Parameter Settings Press in the measurement mode to enter the parameter setting mode P3 0 press to switch the...

Page 15: ...1 Press Aut flashes Press to select auto salinity compensation Aut or manual salinity compensation Hnd Press to confirm 2 Press to enter P3 4 mode or press to return to measurement mode P3 4 Salinity...

Page 16: ...to measurement mode P4 2 temperature unit 1 Press flashes Press to select Press to confirm 2 Press to enter P4 3 mode or to return to measurement P4 3 Auto Lock Off On 1 Press Off flashes Press to se...

Page 17: ...es press or to adjust value then press to confirm 2 Press to enter in P4 8 mode or press to return to measurement P4 8 Adjust time 1 Press HH flashes then press MM flashes When the number flashes pres...

Page 18: ...nload press this key to download the data from the meter to the computer Clear press this key to clear all the stored values Export press this key to export the stored value to a Microsoft Excel file...

Page 19: ...e measuring information is downloaded to the computer through USB and will not be stored in the meter b The stored mode and unit during operation is the same as the data shown on the meter press to ch...

Page 20: ...the warranty A Fails to install operate or use the product in accordance with the instruction manual or if the product is damaged by abuse or incorrect use B The sensor cap is damaged by external forc...

Page 21: ...t can be replaced DO readings are not accurate 1 Check whether the temperature is stable the salinity input and barometric pressure are accurate 2 If the probe calibration is not good recalibrate 3 Ch...

Page 22: ...0 9 67 9 15 8 67 8 21 7 77 7 36 18 0 9 47 8 97 8 50 8 05 7 62 7 22 19 0 9 28 8 79 8 33 7 90 7 48 7 09 20 0 9 09 8 62 8 17 7 75 7 35 6 96 21 0 8 92 8 46 8 02 7 61 7 21 6 84 22 0 8 74 8 30 7 87 7 47 7...

Page 23: ...83 24 83 630 8 84 10 841 00 97 29 02 737 2 98 29 982 85 82 24 54 623 2 83 09 830 87 96 28 72 729 6 97 27 972 72 81 24 24 615 6 82 07 820 73 95 28 43 722 0 96 26 962 59 80 23 94 608 0 81 06 810 60 94 2...

Page 24: ...24 APERA INSTRUMENTS LLC Address 6656 Busch Blvd Columbus Ohio 43229 Tel 1 614 285 3080 Email info aperainst com Website https aperainst com...

Reviews: