3)  For pH Calibration buffer solutions, we recommend that users replace new buffer solution after 

10 to 15 times of use to keep the standard buffer’s accuracy. 

Do NOT pour the used calibration 

solutions back into the solution bottles in case of contamination. 

4)  When calibrating strong basic solutions (pH>10), use 7.00 pH and 12.45 pH to calibrate; When 

measuring strong acidic solutions (pH<4), 7.00 pH and 1.68 pH should be calibrated. 

5)  For the self-diagnosis information, please refer to the table below: 

Symbol  Self-Diagnosis information 

Potential problems and how to fix 

 

 

 
 

The pH calibration solution 
cannot  be  recognized  by 
the meter. 

1. 

Make sure the probe is fully immersed in the 
calibration solution.

 

2. 

Check if calibration solution is expired or polluted.

 

3. 

1st point of pH calibration must be pH 7.00 or 6.86. 
See 6.2 (1).

 

4. 

Please check whether pH electrode is damaged or 
broken. If so, please replace with a new one.

 

5. 

The glass bulb or junction is severely 
contaminated. Please use a soft brush with soap 
water to clean it thoroughly. Then soak it in 3M 
KCL 3-5 hours before performing calibration again.

 

6. 

If none of the above, please contact us at 

[email protected]

 

 

 

 

  Is pressed before 

measurement is fully stable   

Wait for 

  to come up and stay on screen 

before pressing 

 

 

During  calibration,  readings 
being  unstable  for  over  3 
minutes 

1. 

Please check whether pH electrode is damaged or 
broken. If so, please replace with a new one.

 

2. 

The glass bulb or junction is severely 
contaminated. Please use a soft brush with soap 
water to clean it thoroughly. Then soak it in 3M 
KCL overnight before performing calibration again.

 

3. 

The electrode is aged (used for over a year and 
has a much slower response). A replacement is 
needed.

 

4.

 

If none of the above, please contact us at 

[email protected]

 

 

pH  electrode  zero  electric 
potential  out  of  range  (<-
60mV or >60mV) 

1. 

Check whether pH buffer solutions comply with the 
USA or NIST standard.

 

2. 

Check whether pH buffers are expired or 
contaminated.

 

3. 

Please check whether pH electrode is damaged or 
broken. If so, please replace with a new one.

 

4. 

The electrode is aged (used for over a year and 
has a much slower response). A replacement is 
needed.

 

5. 

The electrode is invalidated (Er4/Er5 repetitively 
appears, and problems 1, 2, 3 are excluded). A 
replacement is needed.

 

6. 

If none of the above, please contact us at 

[email protected]

 

 

pH  electrode  slope  out  of 
range 

 

(<85% or >110%)

 

 

The  calibration  reminder  is 
triggered. I

t’s t

ime to perform 

a new pH calibration 

Perform pH calibration or cancel calibration reminder in 
ZenTest settings.

 

Summary of Contents for 60-Z Series

Page 1: ...1 60 Z Series Smart pH Tester pH Temp ORP Instruction Manual PH60 Z pH Tester PH60S Z Spear pH Tester PH60F Z Flat pH Tester www aperainst de V1 0...

Page 2: ...Probe Replacement 11 13 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leave...

Page 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone followin...

Page 4: ...uetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode...

Page 5: ...the storage solution before using b When the tester is not in use we recommend adding one to two drops of tap water to the probe cap and close the cap be careful not to add too much water This way th...

Page 6: ...y rinse the probe in distilled water and dry it and place the probe in 10 01 buffer solution sold separately to complete 3rd point of calibration according to the steps in 6 will appear at the bottom...

Page 7: ...pressed before measurement is fully stable Wait for to come up and stay on screen before pressing During calibration readings being unstable for over 3 minutes 1 Please check whether pH electrode is...

Page 8: ...lid medium When performing such tests pay attention to insert probe evenly and be careful to prevent probe from damage If the medium is too hard such as meat or fruits please bore a small hole with a...

Page 9: ...screw the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14 for how to replace a probe Rinse the probe in distilled wa...

Page 10: ...on will be turned off Long press to manually shut down the meter 5 pH Calibration Reminder P07 set X hours H Or X days D in ZenTest mobile app settings Parameter pH Calibration Reminder On the meter y...

Page 11: ...onsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized...

Reviews: