Apera Instruments 28APE190350/AI31020 User Manual Download Page 11

10 

 

Display similar 

readings in any 

solutions or 

always display 

7.0 pH 

Broken probe 

Make sure the soil is wet. Testing in dry soil will damage the 

probe. If you don’t find any visible damage of the probe and it’s 

within the 1-year probe warranty, contact your point of purchase 

for warranty fulfillment; If there is visible damage or the probe is 

more than 1-year old, replace the probe.

 

Reading keeps 

jumping 

Probe is not fully 

submerged in the 

solution 

Make sure the probe is fully immersed in the solution at least 1 

inch deep 

Air bubbles around 

the sensor 

Make a quick stir in the solution to remove air bubbles 

Probe is not properly 

connected, or the pin 

connector is broken. 

Check the 

probe’s connector, make sure it’s not broken and is 

correctly connected. Align the probe and instrument correctly 

before plugging in. Never force it. Ensure that the probe 

connector is not exposed in the air too long. 

Calibration is 

successful, but 

reading is not 

accurate or 

consistent 

Contaminated probe 

After each measurement, make sure to use a soft brush with 

water to clean off all the dirt on the probe, especially the white 
PTFE junction ring. Avoid wiping or rubbing the blue spear tip. 

Soil difference 

Soil in different locations and different depth will have slightly 

different pH values. Even when the angle at which you stick in 

the probe is different, the measurement can also be affected. 

Therefore, selecting multiple locations and sticking in the probe 

at the same depth and angle, then calculating the average is 

the best way to maximize accuracy. 

Aged probe 

Replace the probe 

Air bubbles around 

the sensor 

Make a quick stir in the solution to remove air bubbles 

Clogged junction 

Clean the probe with cleaning solution, then soak it in 3M KCl 

soaking solution overnight 

Comparison with 

other testers, test 

strips, or drop tests 

To compare with other testers, make sure to perform a 2-point 

calibration for all testers in the same standards, then test a 3rd 

point. Whichever gives more accurate reading in the 3rd point 

standard is the most accurate one. 

Test strips or drop 

tests’ accuracy is not comparable to pH 

meters’. 

Summary of Contents for 28APE190350/AI31020

Page 1: ...GroStar GS2 Premium Soil pH Pen Tester User Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Page 2: ...for the horticultural market Since 1991 Apera instruments has been dedicated to providing hi tech accurate lab grade instruments and sensors GroStar s intelligent design reduces the guesswork so you...

Page 3: ...the backlight Start calibration In calibration mode press to cancel calibration Switch between F and C Swiss LabSen Spear Soil pH Probe Backed by Apera s proprietary sensor technology from Switzerland...

Page 4: ...H Calibration 3 1 Power on the pen and remove the probe cap Always calibrate 7 00 pH first 3 2 Rinse the probe with clean water and shake off excess water then submerge it in the 7 00 pH standard buff...

Page 5: ...f a discrepancy is found then it s time to calibrate again 3 6 If the calibration fails the screen will turn red For details see Section 13 Troubleshooting Guide Always perform at least a 2 point pH c...

Page 6: ...the white PTFE junction ring using the cleaning brush Avoid brushing the spear glass tip After cleaning shake off excess water For details of probe cleaning refer to Section 8 5 6 Repeat Step 5 2 to...

Page 7: ...cap Rinse the probe in the water bucket and shake dry 6 8 Submerge the probe into the soil solution shake for a few seconds and wait for the reading to fully stabilize Then record the reading 6 9 Aft...

Page 8: ...robe in Apera s cleaning solution or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for 1 hour Rinse it off then r...

Page 9: ...w probe make sure to align the connector s position properly and screw on the probe ring pH probes don t last forever Every pH probe will eventually age and fail even if you don t use it that often Th...

Page 10: ...fic instrument manufacturers Contaminated probe Clean the probe with Apera s cleaning solution or detergent water Aged probe Replace the probe Dried out probe Soak the probe in the 3M KCl soaking solu...

Page 11: ...or consistent Contaminated probe After each measurement make sure to use a soft brush with water to clean off all the dirt on the probe especially the white PTFE junction ring Avoid wiping or rubbing...

Page 12: ...10 pH 10 01 solution sold separately Unit pH F C Power supply 4 AAA alkaline batteries Backlight White measurement Green calibration Red errors Reading hold Manual Warranty Two years for the instrumen...

Page 13: ...receipt for Product showing model number payment and date of purchase This warranty is non transferable and terminates if the original purchaser consumer sells or transfers the Product a third party...

Page 14: ...horized distributor or reseller How Do You Get Service To begin a warranty claim you must return the Product to the point of purchase with valid proof of purchase Limitation of Liability Acknowledgmen...

Page 15: ...MAINING PROVISIONS OF THE WARRANTY SHALL REMAIN IN FULL FORCE AND EFFECT Governing Law Authority This warranty is governed by the laws of the state of country where Product is purchased without regard...

Reviews: