B.3.3
Resistance simulation
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Ohm measurement function
Select the Idc measurement function, then the range using the function and navigation keys.
Select the measurement current
Confirm with ENTER.
The following ranges are available:
Range
40 Ohms
400 Ohms
4000 Ohms
Resolution
1 mOhms
1 mOhms
10 mOhms
Measurement current
1-10 mA
0.1-1 mA or
1-10 mA
0.1-1 mA
Settling time
< 1 ms
< 1 ms
< 1 ms
If the measurement current is very high, the message “Out of Range” is displayed in the transmission window.
The resistance simulation function can be used with either 2, 3 or 4 wire connection.
If a polling acquisition system is used, ensure that the current is maintained for more than 1 ms to avoid measurement
errors due to the response time of the resistance simulation function.