Preparation and Inspection
- 7 -
For Veterinary use only
Fill water bottle with purified water to 80% of the capacity, tighten the bottle cap, and connect the water
bottle and the air/water supply connector (inapplicable to VET-3512) of the endoscope. Switch the pump
ON.
Immerse the bending section of the endoscope in a cup with purified water, ensuring the air is fed from
the nozzle at the distal end when cover the air/water valve’s hole (inapplicable to VET-3512), and the
water is fed onto the objective lens when press air/water valve.
Ensure the switches of the imaging processor and the pump are OFF, and then connect the power supply
cords of the imaging processor and the monitor with the wall mains outlet.
Turn on the light; the twinkling LED lamp at the endoscope connector of the endoscope indicates it works
normally.
Turn on the imaging processor and the monitor, check whether the imaging color on the monitor is normal
or not.
3.3.1
Inspection of the insertion section
Visually inspect the surface of the insertion tube for dents, bulges, swelling, fractures or other
irregularities.
Carefully run your hand back and forth over the entire insertion section, ensuring no dents, bulges,
swelling on the insertion section. Also, ensure the insertion section is not slack and no other irregularities.
CAUTION
The power supply switch of imaging processor must be OFF before cutting off the power of the
imaging processor and monitor. Otherwise, the damage caused by it is not warranted by Aohua
in any manner.
3.3 Inspection of the endoscope
CAUTION
The insertion tube surface shall not have any defects that may cause injury to the patients. Stop
using the endoscope even the abnormality found is very tiny, and contact AOHUA.