Setup
To enter the setup mode (Fig. 3) press and hold button 1
for more than a second.
From here use buttons 2 (UP) and 4 (DOWN) to cycle
through the menu screens and button 6 to access the
selected menu screen (Fig. 4).
Buttons 2 and 4 can then be used to alter the alarm
threshold on menu P 1.1 and the temperature scale on
menu P 3.1 (Fig. 4).
Use button 6 to save the current setting and return to the
main menu screen (Fig. 3).
Press button 3 at any time to exit the setup menu.
Fresh Air Correction & Calibration
CO
2
Fresh Air Correction
It is strongly suggested that the unit be corrected in a
sunny, outdoor environment that is well ventilated. Do
not correct the unit in places crowded with people or
close to areas with high CO
2
concentration such as
ventilating outlets or fireplaces.
To correct the unit for CO
2
, place the meter in the
calibration site and press and hold buttons 2 and 3
simultaneously. The unit will flash continually for about
5 minutes, then return to normal operation.
To abort correction at any time, simply turn the meter
off.
CAUTION:
The meter is corrected at ambient air of
around 400ppm. Do not correct the meter in an
environment with an unknown CO
2
concentration.
Doing this could lead to inaccurate measurements.
Specification
•
CO
2
Range of 0 -9999
ppm
(maximum alarm threshold 9900
ppm
)
Resolution of 2
ppm
Accuracy ± 75
ppm
±5%
•
Temperature
Range of -10 º to 60 ºC (14 -140 ºF)
Resolution of 0.1 ºC (0.1 ºF)
Accuracy ± 0.6 ºC
•
Warm Up
30 seconds
•
Operating Ranges
From 0~50ºC, 0 ~ 95% RH (avoid
condensation)
•
Storage
From -20 º ~ 60 ºC, 0 ~ 99% RH (avoid
condensation)
•
Power
4 pcs AA batteries, DC Adaptor
•
Battery Life
10 hours (Alkaline)
•
Warranty Period
3 years (anticipated sensor life 15 years)
Troubleshooting
•
Fails to power up
Press button 1 for more than 0.3 seconds
and try again. Check whether batteries are
in good contact and correct polarity is
observed.
•
Fixed readings
Check whether data hold function was
activated.
•
Slow response
Check whether the air flow channels on the
rear are blocked.
•
Error messages
E01: CO
2
sensor damaged
E02:The value is under range
E03:The value is over range
E04:The original data error results in this
error
E07: Voltage to measure CO
2
. Replace
batteries or use an adaptor
E17: Retry CO
2
calibration
E31: Temperature sensor damaged