Setup  

To enter the setup mode (Fig. 3) press and hold button 1 
for more than a second.  

From  here  use  buttons  2  (UP)  and  4  (DOWN)  to  cycle 
through  the  menu  screens  and  button  6  to  access  the 
selected menu screen (Fig. 4).  

Buttons  2  and  4  can  then  be  used  to  alter  the  alarm 
threshold  on  menu  P  1.1  and  the  temperature  scale  on 
menu P 3.1 (Fig. 4).  

Use button 6 to save the current setting and return to the 
main menu screen (Fig. 3).  

Press button 3 at any time to exit the setup menu.  

   

Fresh Air Correction & Calibration  

CO

Fresh Air Correction 

 

It is strongly suggested that the unit be corrected in a 
sunny, outdoor environment that is well ventilated. Do 
not  correct  the  unit  in  places  crowded  with  people  or 
close  to  areas  with  high  CO

concentration  such  as 

ventilating outlets or fireplaces.  

To  correct  the  unit  for  CO

2

,  place  the  meter  in  the 

calibration  site  and  press  and  hold  buttons  2  and  3 
simultaneously. The unit will flash continually for about 
5 minutes, then return to normal operation.  

To  abort  correction  at  any  time,  simply  turn  the  meter 
off.  

CAUTION: 

The  meter  is  corrected  at  ambient  air  of 

around  400ppm.  Do  not  correct  the  meter  in  an 
environment  with  an  unknown  CO

concentration. 

Doing this could lead to inaccurate measurements.  

 

   
   

 
 
 
 
 
 
 

   

   
   
   

   
   
   

   
   

   
   

 

Specification  

 

CO

 

Range of 0 -9999

ppm  

(maximum alarm threshold 9900

ppm

Resolution of 2

ppm 

Accuracy ± 75

ppm 

 ±5% 

 

Temperature 

 

Range of -10 º to 60 ºC (14 -140 ºF) 
Resolution of 0.1 ºC (0.1 ºF) 
Accuracy ± 0.6 ºC  

 

Warm Up 

 

30 seconds  

 

Operating Ranges 

 

From 0~50ºC, 0 ~ 95% RH (avoid 
condensation)  

 

Storage 

 

From -20 º ~ 60 ºC, 0 ~ 99% RH (avoid 
condensation)  

 

Power 

 

4 pcs AA batteries, DC Adaptor

 

 

Battery Life 

 

10 hours (Alkaline)  

 

Warranty Period 

3 years (anticipated sensor life 15 years) 
   

Troubleshooting  

 

Fails to power up 

 

Press button 1 for more than 0.3 seconds 
and try again. Check whether batteries are 
in good contact and correct polarity is 
observed.  

 

Fixed readings 

 

Check whether data hold function was 
activated.  

 

Slow response 

 

Check whether the air flow channels on the 
rear are blocked.  

 

Error messages 

 

E01: CO

sensor damaged 

E02:The value is under range 
E03:The value is over range 
E04:The original data error results in this     
error 
E07: Voltage to measure CO

2

. Replace 

batteries or use an adaptor 
E17: Retry CO

calibration 

E31:   Temperature   sensor   damaged   

   

   

   