4 Installation
22
D52IB001EN-G
Fig. 25:
Installation diagram for TTK 600 with compressed-air cooling equipment.
4.2 Installation Requirements
The setup location and surroundings should meet
the requirements of a typical laboratory.
Allow the equipment to reach ambient temperature
before installation. This is very important if the
equipment has been stored or transported at lower
temperatures.
To ensure temperature stability and trouble-free
measurement never locate your instrument:
• next to a heating facility
• near an air conditioning, ventilation system or an
open window
• in direct sunlight
Keep the instrument away from magnetic fields.
Read the Safety Instructions in chapter 1.
Find all Technical Data in appendix A.
4.2.1 X-Ray Diffractometer Requirements
• Radiation safety enclosure
• Vertical
θ
-
θ
goniometer
Operation on
θ
-2
θ
goniometer is possible, but
during low temperature operation the LN2 hoses
get stiff and rotatability of the sample chamber
may be limited.
• Clearance around goniometer axis:
at least
125 mm
=> see also appendix A
• Feedthroughs for two LN2 hoses with
40 mm
each, two water hoses with 4x6 mm, two electri-
cal cables with
5 mm and a vacuum hose with
CCU 100
rear panel
POWER
RS 232
AIR
OUT
1
AIR
IN
LAN
Sensors
Heater
Heater
Out 1
Out 2
Flow
Sensor
Low Power
Module
TTK 600
SAMPLE
CHAMBER
PC
compr
essed air
⇒
L=5m
L=5m
water out
⇒
P
⇐
water in
DIFFRACTOMETER CABINET
4
3
WATER FLOW
MONITOR
PRESSURE
REGULATOR
L=3m
air out
2
2 bar
max. 8 bar
AIR DRYER
(optional)
5
AIR SERVICE
UNIT
UNIT
P
⇐
compressed-air in
max. 8 bar
min. 6 bar
max. 8 bar
WARNING
Risk of injury.
Operation of TTK 600 Low-Temperature Cham-
ber on an X-ray diffractometer without radiation
safety enclosure can cause exposure to ionizing
radiation.
Exposure to ionizing radiation above natural
dose can lead to skin and eye damage and can
have carcinogenic and mutagenic effects in your
body.