7.3.1.
Disinfection cycle with oxygenated water
With the S.H.S. system, you can execute a manual disinfection cycle of
the water ducts of all the instruments on the dentist’s instrument board
and the syringe on the assistant’s board using oxygenated water (hydro-
gen peroxide).
To disinfect, proceed as directed below:
A)
Prepare the disinfectant:
• Pour undiluted PEROXY Ag+ (not available for the American and
Canadian markets) (or 3% oxygenated water) into the tank marked
with an orange band.
NOTE:
make sure that the tank is completely filled.
B)
Putting in the disinfectant:
• Check that the spray tap of each dynamic instrument is completely
open (if not, no or too little water will come out).
• Make sure the spray cocks (
c
) found towards the bottom of the
board, are open (if they are not, either very little or no water at all
will flow out).
• Take out the bottle of distilled water (
a
) turning it clockwise.
• Replace the bottle of distilled water (
a
) with the one containing
disinfectant.
•
Remove the syringe and empty the duct with the water button
.
NOTE:
when emptying is complete, no water will come out
of the duct.
•
Repeat the same procedure for the syringe on the assistant’s board
(with the water key). Use the Water clean system with the foot control
for the remaining dynamic instruments (see paragraph 5.2)
.
• Turn the toggle (
c
) to position "OPEN AIR PRESSURE”.
•
Remove the instruments and fill the ducts with the hydrogen peroxide
solution. Use the Water clean system with the foot control for the
dynamic instruments and the water button for the syringes.
NOTE:
let the disinfectant flow from the instruments for about
6-7 seconds.
• Put the instruments back in place.
NOTE:
at this point, the ducts contain the disinfectant.
C)
Putting in disinfectant
WARNING!
The disinfectant must be left to dwell in the ducts for at least 10
minutes, but not more than 30 minutes.
D)
Rinsing the ducts:
• Turn the toggle (
c
) to position "CLOSE AIR PRESSURE”.
• Unscrew the bottle (
a
) containing the disinfectant turning it clockwise.
• Put in the bottle containing distilled water.
• Remove the syringe and empty the water duct.
NOTE:
when emptying is complete, no water will come out of the duct.
• Repeat the same procedure to drain the hydrogen peroxide from the ducts for the various instruments.
• Turn the toggle (
c
) to position "OPEN AIR PRESSURE”.
• Remove the instruments and wait for distilled water to flow out. Use the Water clean system (see paragraph 5.2) with the foot control for the
dynamic instruments and the water button for the syringes.
NOTE:
allow water to flow out approximately 6-7 seconds.
•
At this point, the ducts contain distilled water again and the dental unit is ready to be used.
WARNING!
• After disinfection, make sure that you close the tank containing the disinfectant (exposed to air, it loses its effectiveness).
•
It is good practice to perform a disinfection cycle at least once a day, preferably at the end of the day.
PEROXY Ag+ storage.
For proper storage of
PEROXY Ag+
follow the manufacturer’s instructions printed on the package.
It is important to keep the package tightly closed and store it in a cool place at a temperature not exceeding 25°C.
WARNING!
Never leave PEROXY Ag+ or oxygenated water in the tank marked with an orange band for more than one month.
In case of long absences from the studio (holidays), completely empty out the tank marked with an orange band before leaving.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......