40
40
EN
R7
- OPERATING INSTRUCTION
5.3. Syringe
Description of the instrument.
[ a ]
Nozzle.
[ b ]
Handpiece.
[ c ]
Syringe release button.
[ d ]
Air button.
[ e ]
Water button.
[ f ]
Hot/cold selector.
[ g ]
Hot/cold indicator light.
Technical charachteristics.
• Operating time:
- 3F syringe: continuous operation,
- 6F syringe: 5 sec. operation, 10 sec. rest.
• Power supply:
- 6F syringe (CEFLA s.c. models): 24 Vac; 50/60 Hz; 2 A; 50 W.
•
Classification in accordance with standard EN 60601-1:
- 6F syringe (CEFLA s.c. models): CLASS II, type B.
• Installation plan: consult the Technical Installation Manual (see Paragraph 11.).
Operation.
• Place the instrument in its work position.
• Button (
e
) = water;
Button (
d
) = air;
Buttons (
e + d
) = spray.
• 6F syringe, operation with hot water, air and spray: turn the selector [ f ]
clockwise (LED g on).
• 6F syringe, operation with cold water, air and spray: turn the selector
[ f ] anticlockwise (LED g off).
Removing the handpiece.
• The nozzle (
a
) is screwed onto the grip (
b
).
• Turn the selector switch counter-clockwise (LED
g
off) and press the
button (
c
) to take the grip off the syringe casing.
Removable syringe cord
The syringe has a removable cord to ease cleaning (see chapter 5).
Cleaning.
Use soft disposable paper towel dampened with detergents/disinfectants
WARNING!
• Do not soak the syringe in liquid disinfectants or detergents.
• Products not recommended: harsh products and/or products con-
taining acetone, chlorine and sodium hypochlorites.
Sterilization.
Syringe grip and spout: see paragraph 1.5.
NOTE:
Bag before sterilizing.
Instructions for use.
WARNING!
• The instrument is supplied NOT STERILE and must be sterilized
before use (see paragraph 1.5.).
• It is recommended to use single-use protections and nozzles.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......