Anthogyr MICRO NITI 10608 Instructions For Use Manual Download Page 4

12 

Install the MD  on the motor connection until ratcheting. For this, 
keep the motor and the MD in the same axle. The motor must be at 
a complete standstill. 
Apply light traction to the MD to check that it is properly attached to 
the motor connection. 
Operational test by switching on the motor; start at low speed, then 
gradually increase, outside of the patient’s mouth. 
If you notice overheating, irregularities, vibrations, abnormal noises 
when using the MD, immediately contact your AS. 

5.2 – Adjusting the declutching ring (Fig. 3) 

The torque control system cannot in any way be a substitute for the 
practitioner’s assessment and experience. 
Increasing the declutching threshold : 
• Push ring G towards Head E on the MD, notch by notch, until the 
right position is reached. 
Decreasing the declutching threshold : 
• Pull ring G towards the body H on the MD, notch by notch, until 
the right position is reached. 
To  adjust  the  declutching  threshold  (position  1,  2,  3,  4),  start  by 
setting it at its minimum torque (position 1). 
If the MD keeps declutching, move up to a higher torque (Positions 
2 to 4). 
When the radicular instrument stops, a slight clicking sound may 
be heard; in that case, lower the pressure on the MD
In the event that the instrument becomes wedged in the root canal, 
reverse direction of the motor rotation. 

5.3 – Disconnection of motor 

Remove the MD by keeping it in the motor axle. 

  In the event of prolonged non use of the device, do not leave 
  it connected to the motor. Risk of damaging the motor. 

5.4 – Connection/disconnection of rotary instrument 

  For  all  handling  of  rotary  instruments,  wear  protective 
  gloves. Risk of cutting and contamination. The tool must only 

be handled when the motor is at a standstill. 
Only use rotary instruments in good condition. Risk of necrosis. 
Insertion of rotary instrument 

Contra angle with push button : 
• Insert the rotary instrument into the head of the MD  until it hits 
the stop. 
• Turn the rotary instument whilst maintaining a slight axial pres-
sure until it clicks (Fig. 4). 

Contra angle with latch : 
• Open the latch. 
• Insert the rotary instrument into the head of the MD  until it hits 
the stop. 
• Turn the rotary instument whilst maintaining a slight axial pres-
sure until it reaches the running slot on the drive plate. 
• Lock the latch (Fig. 5). 

  For tool replacement: check the good condition of the rotary 
  tool with light axial traction. 

MD 

lifecycle  

If  used  in  a  proper  manner,  all  MD  parts  have  a  lifecycle corre-
sponding to 250 sterilisation cycles. 
However, these indications are not a warranty because wear may 
appear  prematurely,  depending  on  how  the  MD  is  maintained 
(cleaning and sterilisation). 
 

VI. HYGIENE AND MAINTENANCE 

  Sterilisation  of  medical  devices  must  be  done  by  properly 
  trained  and  protected  staff,  in  compliance  with  current 

regulations. The sterilisation protocol must be appropriate to  the 
infectious risk. 
Wear appropriate protective clothing: risk of infection and injury. 
Only use products for the maintenance of medico-surgical equip-
ment  compatible  with  aluminium.  Detergents  and  alcaline  type 

F

ra

n

ça

is

 

En

g

li

sh 

(U

K

)

 

D

e

u

tsch

 

Esp

a

ñ

o

It

al

iano

 

P

o

rt

u

g

u

ê

s

 

N

e

d

e

rla

n

d

s

 

中国的

 

13 

   

  ر

ع

 

En

g

li

sh 

(U

S

)

 

solutions (pH > 7) damage aluminium based instruments. Forbid 
antiseptics that are intended for use on skin and mucous. 
Forbid  products  containing  aldehyde,  alcohol  or  other  products 
likely to bind proteins. 

  For each product used: refer to the manufacturer’s instruc-
  tions. 

Comply with the concentrations, exposure durations, and life span 
of products. Do not mix the products and follow the instructions 
for their disposal. 

6.1 – Preparation of 

MD

 in the place of use 

To be done immediately after surgery (maximum time of 2 hrs): 
• Disconnect the MD from the motor, 
• Remove the rotary tool (see § 5.6), 
•  Remove  the  largest  organic  contaminants  with  disinfectant 
wipes. 

6.2 – Preparation for sterilisation 

• Disconnect the MD from the motor, 
• Remove the rotary instrument. 

6.2.1 – Manual preparation for sterilisation 

  By brushing: 
  • Brush each part under running water with a soft 

brush. 
• Wipe a disinfectant wipe over each of the MD’s subunits, 
• Or Spray disinfectant on the MD and wipe with a clean cloth. 

Rinsing and drying: 
• Rinse well

2

,then dry each part. 

6.2.2 - Automatic preparation for sterilisation 

  • Only in a washing and disinfecting machine, appropriate 
  for the treatment of this type of MD. Follow the manufac-

turer’s instructions for the device. 
• Place the assembled MD on the nozzle for MD
• The disinfection thermo cycle must be at least 10 minutes at 93°
C (203°F). 
• Check there is no residue and that all the parts of the MD  are 
completely dry at the end of the cycle. 

  • If the washing and disinfecting machine is near the place 
  of use, it is possible to also do step 6.1. 

6.3 - Lubrication 

  Follow the spray lubricant manufacturer’s instructions. 
  Wear appropriate protective clothing (splashes, etc.). 

Keep away from all sources of heat or ignition. In particular: do not 
smoke. Risk of inflammation. 
After cleaning and before each sterilisation (once a day minimum): 
• Remove the rotary instrument, 
• Insert the tip to the back of the MD (Fig 6), 
• Cover the head of the MD with a soft cloth or paper or a wipe, 
• Point the head towards the bottom, 
• Spray several times by firmly holding the instrument, 
• Wipe excess oil with a cloth or a wipe. 

6.4 – Operational test 

• Connect the MD  to a micro-motor, point the head towards the 
bottom. 
• Turn the micro-motor on at low speed for 30s. Gradually bring 
up to maximum speed. 
• Wipe excess oil with a cloth or a wipe 

  If  you  notice  overheating,  irregularities,  vibrations,  abnor-
  mal noises when using the MD, immediately contact your 

AS. 

6.5 - Sterilisation 

  • The instruments are to be sterilised before first use and 
  after each use. 

• Only sterilise cleaned, lubricated and tested instruments. 
•  We  recommend  sterilisation  that  complies  with  the  EN  13060 
standard, class B. All other methods are forbidden. 
• Independently  put each MD  in a sterilisation sachet that com-
plies with current standards and the size of the MD
• Comply with the space between the sachets and do not overload 
the autoclave. 

: Temperature  <  38°C,  distilled water is recommended instead  of running  water  if  this has too 

much chlorine (cf. FD98-135 standard). 