Anritsu VectorStar MN4690B Series Quick Start Manual Download Page 4

QSG-4

PN: 10410-00290  Rev. A

MN4690A QSG

4.

GPIB Address Settings

The MN4690B Series Test Set GPIB address must match the GPIB address set on the VNA and is set on the 
Test Set by rear panel Dual Inline Package (DIP) switches. The factory default GPIB address is 

16

 (where 

Switch 1 = ON or up and all other switches OFF or down). The GPIB address on the VNA can be verified by 
navigating to the 

REMOTE INTER

 menu and the 

Multiport Test Set

 button.

Navigation: 

MAIN | System | SYSTEM | Remote Interface | REMOTE INTER | Multiport Test Set

  

5.

Power Up Sequence

1. 

Make sure all of the cables are properly installed as illustrated in the previous sections.

2. 

Turn on the Test Set prior to launching the VNA application on the VNA. The VNA can be powered on, 
but the VectorStar application must be off.

3. 

Once the Test Set is on, launch the VNA application. During the launch sequence, the VNA application 
recognizes the Test Set on the GPIB bus and is configured for 4-port mode. If the VNA application is 
powered up and launched before the test set, the VNA application will stay in 2-port mode and only 
2-port mode features and functions will be available.

4. 

If the 4-port functions fail to appear, exit the VNA application by selecting from the VectorStar Menu 
Bar: 

File

 | 

Exit

 and clicking 

Yes

 in the resulting dialog box.

5. 

After the Windows desktop appears, launch the VNA application by doing one of the following:

a.

On the desktop, click the 

VectorStar

 icon. If the VNA was running in 100,000 point mode, the icon 

is annotated with “100K”.

b.

If running in 25,000 point mode, select 

Start

 | All Programs | VectorStar | VectorStar

.

c.

If running in 100,000 point mode, select 

Start

 | All Programs | VectorStar_100K | VectorStar.

Figure 4.

MN4690B DIP Switches (Factory Default Setting)

Note

Note that the VNA GPIB connection must be made to the 

VNA Dedicated GPIB 

connector and 

NOT

 

to the 

VNA IEEE-488.2 GPIB

 connector. 

Note

The VNA application must be started 

AFTER

 the Test Set is connected and powered up. If the VNA 

application is started before the Test Set, the VNA will remain in 2-port mode and the 4-port functions 
will not be available. If this happens, exit the VNA application, make sure the Test Set is powered up, 
and launch the VNA application.