3-6 Directivity and Test Port Match Verification
Chapter 3 — Performance Verification
3-16
PN: 10410-00730 Rev. D
MN469xC Series Multiport Test Set MM
•
For ripple with a negative trough:
Directivity = Return Loss value
|Mkr 2 value|
|Ref
X value|
•
For ripple with a positive peak:
Directivity = Return Loss value
|Mkr 2 value|
|Ref
X value|
•
Example:
Directivity = 20 dB + 17.452 dB - 0.9151 dB = 36.5369 dB
84.
Record the directivity value into the Port 1 Measured column of the following applicable table in
:
•
Table A-1, “Directivity Record for MN4694C Multiport VNA System” on page A-2
for MN4694C
•
Table A-3, “Directivity Record for MN4697C Multiport VNA System” on page A-4
for MN4697C.
85.
Repeat Step
to Step
for other frequency bands listed in
MN4697C in
86.
Disconnect the Offset Termination from the Air Line and connect a short.
87.
Select
Scale
|
Auto Scale Active Trace
88.
Repeat Step
to Step
. Record the Return Loss value into the appropriate table and the Port 1
Measured column in
•
Table A-2, “Test Port Match Record for MN4694C Multiport VNA System” on page A-3
MN4694C
•
Table A-4, “Test Port Match Record for MN4697C Multiport VNA System” on page A-5
MN4697C.
89.
Repeat Step
for other frequency bands listed in
for MN4694C or
for MN4697C in
.
90.
Disconnect the Short from the Air Line and then disconnect the Air Line from the Test Port.
91.
Select
Trace
|
Trace Next
92.
Connect the Air Line to the next Test Port (e.g. Port 2 for S22, Port 3 for S33 or Port 4 for S44) on
MN469xC and then terminate the Air Line with the Offset Termination.
93.
to verify the Directivity and Port Match of the Test Port being tested.
94.
for the rest of the Test Ports on MN469xC.