3-6 Directivity and Test Port Match Verification
Chapter 3 — Performance Verification
3-12
PN: 10410-00730 Rev. D
MN469xC Series Multiport Test Set MM
39.
Select
Port 1 Reflective Device
to display the Refl. Device(s) Menu. See
below.
40.
Connect the Open from the Calibration Kit to Port 1 of the VNA.
41.
Select
Open
.
42.
Disconnect the Open from Port 1 and connect the Short.
43.
Select
Short
44.
Disconnect the Short from Port 1 and connect the termination.
45.
Select
Load
46.
Select
Back
to return to previous menu.
Figure 3-5.
Refl. Device(s) Menu
Note
Use the Anritsu 01-201 torque wrench from the Anritsu 365xx Calibration Kit to tighten the
connection to improve the quality of the calibration. Do the same for other calibration standards
(e.g. Short and Load) in subsequent steps.
Note
A check mark will appear on the button when the measurement or process is complete. See
examples on right side of
above.