2-15 Adjacent Channel Power Measurement
Spectrum Analyzer
2-20
PN: 10580-00349 Rev. H
Spectrum Analyzer MG
2-15 Adjacent Channel Power Measurement
Required Equipment
•
30 dB, 50 watt, Bi-Directional, DC–18 GHz, N(m)–N(f) Attenuator, Anritsu
Part Number 42N50A-30 (if required for the power level being measured)
•
Test Port extension cable, Anritsu Part Number 15NNF50-1.5C
Procedure
1.
Using the test port extension cable and 30 dB attenuator, connect the signal source to
the input of the attenuator, and connect the output of the attenuator to the
RF In
test
port.
2.
If an attenuator was connected in step 1, press the
Amplitude
main menu key then
press the
RL Offset
submenu key, enter
30
then select dB
External Loss
to compensate
for the loss in the attenuator.
3.
Press the
Amplitude
main menu key and press the
Reference Level
submenu key to set
the reference level to
60 dBm
.
4.
Press the
Atten Lvl
submenu key to set the input attenuation level needed for the
measurement. This value depends on the input power level and any external
attenuator. Enter an attenuation level to achieve roughly –40 dBm at the input mixer.
5.
Press the
BW
main menu key and verify that
Auto RBW
and
Auto VBW
are
On
.
6.
There are two ways to set the measurement parameters. If the signal standard and
channel are known, press the
Freq
main menu key and set the signal standard and
press
Channel
submenu key for the signal to be measured, then skip to
. If the
signal standard and channel are not known, follow the procedure in
through
.
7.
Press the
Freq
main menu key, press the
Center Freq
submenu key, and enter the
desired center frequency.
8.
Press the
Shift
key then the
Measure
(
4
) key. Then press the
Power and Bandwidth
submenu key, and press the
ACPR
submenu key.
9.
Press the
Main Ch BW
submenu key, and enter the main channel bandwidth.
10.
Press the
Adj Ch BW
submenu key, and enter the adjacent channel bandwidth.
11.
Press the
Ch Spacing
submenu key, and enter the channel spacing.
12.
Make the measurement by pressing the
On/Off
submenu key. The detection method is
automatically changed to RMS Average.
Solid vertical lines are drawn on the display to indicate the main channel. Dashed vertical
lines define the adjacent channels. The SPA will display the measurement results in the
message area.
Note
Adjacent Channel Power Ratio is a constant measurement. After it is turned on, it
will remain on until it is turned off by pressing the On/Off submenu key again.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com
Summary of Contents for Spectrum Master MS2720T
Page 2: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 5: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 22: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 241: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 264: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 265: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 558: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 800: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1027: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1033: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1034: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1035: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...