General Information
1-3 Calibration and Verification
Site Master User Guide
PN: 10580-00252 Rev. K
1-3
Available Options
Available options for the Site Master are shown in
.
1. Requires External Power Sensor.
2. Requires Option 31.
1-3
Calibration and Verification
Anritsu recommends annual calibration and performance verification by local Anritsu
service centers. The Cable and Antenna Analyzer mode requires calibration standards for
OPEN, SHORT, and LOAD (OSL) or InstaCal module, which are sold separately.
The Site Master is self-calibrating and there are no field-adjustable components. The OSL
calibration components are crucial to the integrity of the calibration. As a result, they must be
verified periodically to ensure performance conformity. This is especially important if the
OSL calibration components have been accidentally dropped or over-torqued.
Contact information for Anritsu Service Centers is available at:
http://www.anritsu.com/contact-us
Table 1-2.
Available Options
S331E
S332E
S361E
S362E
Description
S331E-0010
S332E-0010
S361E-0010
S362E-0010
Bias-Tee (Requires Option 21 on
S331E and S361E)
S331E-0019
S332E-0019
S361E-0019
S362E-0019
High-Accuracy Power Meter
(1)
S331E-0021
S332E-0021
S361E-0021
S362E-0021
2-Port Transmission Measurement
S332E-0025
S362E-0025
Interference Analyzer
(2)
S332E-0027
S362E-0027
Channel Scanner
S332E-0028
S362E-0028
C/W Signal Generator (Requires
CW Signal Generator Kit)
S332E-0029
S362E-0029
Power Meter
S331E-0031
S332E-0031
S361E-0031
S362E-0031
GPS Receiver (Requires Anritsu GPS
Antenna)
S332E-0090
S362E-0090
Gated Sweep
S331E-0098
S332E-0098
S361E-0098
S362E-0098
Standard Calibration to
(ANSI 2540-1-1994)
S331E-0099
S332E-0099
S361E-0099
S362E-0099
Premium Calibration to Z540 plus test
data
S331E-0411
S332E-0411
S361E-0411
S362E-0411
Ethernet Connectivity
S332E-0431
S362E-0431
Coverage Mapping
(2)
S332E-0509
S362E-0509
AM/FM/PM Analyzer
S331E-0751
S332E-0751
S361E-0751
S362E-0751
CPRI RF Measurements
Note
Anritsu recommends allowing the instrument to warm up to typical operation
temperature (approximately 15 minutes) before calibrating.
Summary of Contents for S331E
Page 2: ......
Page 36: ...2 9 Tilt Bail Stand Instrument Overview 2 22 PN 10580 00252 Rev K Site Master User Guide ...
Page 48: ...3 4 Spectrum Analyzer Quick Start Guide 3 12 PN 10580 00252 Rev K Site Master User Guide ...
Page 82: ...7 1 Overview Bias Tee Option 10 7 2 PN 10580 00252 Rev K Site Master User Guide ...
Page 100: ...A 1 Introduction Measurement Guides A 2 PN 10580 00252 Rev K Site Master User Guide ...
Page 124: ...Index 4 PN 10580 00252 Rev K Site Master User Guide ...
Page 125: ......