Chapter 5 Remote Control
5-70
:CALCulate:DATA:EALarm? <result>
Parameter
<result>=<STRING PROGRAM DATA>
Refer to Table 5.9-4 for details of <result>.
Response
<string>=<STRING RESPONSE DATA>
Table 5.9-4 Parameter
Items
<result1>
Format
Error Count
"EC"
Form1
Bit Count
"BITS"
Form1
Bit Error Rate
"BER"
Form2
Pass/fail judgment
"JUDGe"
String("PASS","FAIL","---")
Table 5.9-5 Response Format
Items
Format
Description
Form1
Integer
"XXXXXXX"
For 0 to 9999999
"X.XXXXEXX"
For 1.0000E07 to 9.9999E17
"-------"
No data corresponds to a query.
Form2
Decimal
"X.XXXXE-XX"
For 0.0001E–18 to 1.0000E00
"-------"
No data corresponds to a query.
Function
Queries the BER Measurement results (BITS).
Example
> :CALCulate:DATA:EALarm? "BITS"
< "1.0000E12"
:SENSe:MEASure:BER:IPADdress <address>
Parameter
<address>=<STRING PROGRAM DATA>
“xxx.xxx.xxx.xxx”
IP address
Function
Sets an IP address of the USB3.1 Receiver Test Adapter to be connected.
Example
To set the IP address 192.168.0.10:
>:SENSe:MEASure:BER:IPADdress “192.168.0.10”
:SENSe:MEASure:BER:IPADdress?
Response
<address>=<STRING PROGRAM DATA>
“xxx.xxx.xxx.xxx”
Outputs the format 223.255.255.254.
Function
Queries the setting value of the connection destination IP address.
Example
>:SENSe:MEASure:BER:IPADdress?
< “192.168.0.10”
Summary of Contents for MX183000A
Page 12: ...xii ...
Page 16: ...IV ...
Page 20: ...Chapter 1 Outline 1 4 Figure 1 1 3 Jitter Tolerance Graph Screen ...
Page 54: ...Chapter 3 Connecting Equipment 3 16 ...
Page 136: ...Chapter 5 Remote Control 5 30 Figure 5 8 2 Sequence Screen BER Measurement 10 8 9 11 ...
Page 206: ...Chapter 5 Remote Control 5 100 ...
Page 218: ...Appendix A Specifications A 12 ...
Page 226: ...Appendix B Default Settings B 8 ...