4-11 Touch Screen Accuracy Adjustment
Chapter 4 — Adjustment
4-10
PN: 10370-10372 Rev. B
MG3702xA MM
Note this power level and then decrease the DAC value until the power level is 1 dB below the
noted maximum power.
15.
Press
Next.
16.
Adjust the DAC value so the power meter reads 10 dB below the desired level.
17.
Press
Next.
18.
The adjustment may require another iteration. Press
Next
to continue and repeat
if another iteration is required.
19.
Once
Down Converter Detector
is complete touch
Next
to start
Main Band Detector Cal
.
20.
through
to complete
Main Band Detector Cal
.
21.
When the adjustment is complete, press
Next
to continue.
22.
Press
Save to Nonvolatile.
23.
Choose
Yes
to overwrite.
24.
Choose the tab called
Log Amp Zero.
25.
Press
Run Cal | Next.
26.
When adjustment is finished press
Next
27.
Press
Save to Nonvolatile
28.
Choose
Yes
to overwrite.
29.
The calibration is complete. Record as completed in the test record.
4-11 Touch Screen Accuracy Adjustment
This process should be performed when the front panel or A2 Microprocessor PCB has been replaced. The
adjustment improves the accuracy of presses on the touch screen as compared to the mouse pointer.
1.
Enter the display configuration menu (
) by pressing
System
| Confi. | Display…
Figure 4-4.
Display Configuration Menu
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......